Safety and Related Information
Guidance and Manufacturers Declaration for Electromagnetic Immunity
The imager is intended for use in the electromagnetic environment specified below. The customer or user
should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic Environment—Guidance
Electrostatic
discharge (ESD):
EN61000-4-2
•IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst:
EN61000-4-4
•IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge:
EN61000-4-5
•IEC
61000-4-5
± 1 kV
differential mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital environment.