Safety and Related Information
CAUTION:
Double pole/neutral fusing.
CAUTION:
Do not use a cell phone within 2 m of this device. This
proximity includes any imager behind a wall adjacent to your
location.
CAUTION:
Do not use a microwave oven within 4 m of this device.
Electromagnetic radiation from a microwave oven is only an
issue if after the oven door is closed and latched, the seal does
not maintain an electromagnetic tight fit between the oven
door and oven main housing. Determining if the seal has an
electromagnetic tight fit requires special detection equipment.
CAUTION:
Do not use in the presence of flammable anesthetics, oxygen,
or nitrous oxide. This equipment does not have a gas-sealed
electronics enclosure and could ignite any flammable or
explosive gases present in its environment.
CAUTION:
This equipment uses a DICOM network port, and is intended
to connect to other medical devices. It is not intended to be
connected directly outside the building. Only qualified
personnel may perform installation and service.
CAUTION:
This device is NOT to be located within the patient
environment. Therefore, the equipment must not be located
closer than 1.83 m from a patient bed or chair.
CAUTION:
This device can affect implantable pacemaker or implantable
cardioverter defibrillators at distances of less than 30.48 cm
(1.00 ft).
1.83 m