User's Manual

12 9G3888_en 2011-03-17
Safety and Related Information
Guidance and Manufacturer’s Declaration for Electromagnetic Immunity
The Laser Imager is intended for use in the electromagnetic environment specified below. The
customer or user of the Laser Imager should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic Environment -
Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Laser Imager, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz–80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
3 v/m
80 MHz–2.5 GHz
3 v/m 80 MHz–800 MHz
800 MHz–2.5 GHz
d is the recommended separation distance
in meters (m).
P is the maximum output rating of the
transmitter in watts (W) according to the
transmitter manufacturer.
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency
range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
See NOTE 1 on the top of the next page.
b
See NOTE 2 on the top of the next page.
d 1.17 P=
d 1.17 P=