User's Manual
Table Of Contents
- 1 Overview
- Major internal assemblies
- How the laser imager works
- Print sequence
- Film sizes
- Automatic image quality and processing
- Configuring and monitoring the system (using the Web Portal)
- Enhancing serviceability with remote monitoring
- Agency compliance
- User guide conventions
- 2 Basic Operating Tasks
- Understanding the display screen
- Turning the power on and off
- Emergency shutdown or power loss
- Restarting the laser imager
- Working with film cartridges
- Checking film count
- Checking the size of the loaded film
- Film count is flashing "0"
- Inserting a new film cartridge
- Loading a different film size to match a print request
- Deleting pending jobs
- Making a test print
- Calibrating the laser imager for the installed film
- Opening or removing a cover
- Using the Web Portal to access additional functionality
- Accessing the Web Portal
- 3 Maintenance and Troubleshooting
- Overview: Status and error messages and codes
- Preventive maintenance
- Replacing the filter
- 550 code and Maintenance symbol
- Error indicators on the display screen
- Calibration error
- Required restart
- Film jams
- Using the Web Portal to gain more information on errors
- Understanding the codes on the Web Portal and the Display Screen
- Subsystem error codes and messages
- DICOM (Digital Imaging and Communications in Medicine)
- Printer
- Film Cartridge
- Job Manager
- Condition codes
- Correcting film jams
- Film jam code 116 / Jam in Area 1
- Film jam code 323 / Jam in Area 2
- Film jam code 324 or 325 / Jam in Area 2
- Film jam code 326 / Jam in Area 2 or 3
- Film jam code 543 / Jam in Area 3
- Film jam code 544 / Jam in Area 3
- Display Screen is not functional
- Calling for support
- 4 Film Technical Information
- General description
- Spectral sensitivity
- Image quality
- Environmental impact
- Storing and handling undeveloped film
- Dissipating odor
- Recycling film
- 5 Specifications
- Equipment specifications
- Operating space requirements
- Environmental requirements
- Temperature
- Relative humidity
- Altitude
- Surface levelness
- Environmental effects
- Power requirements
- Network requirements
- Publication History
Maintenance and Troubleshooting
2011-03-31 9G3886_en 3-13
607 27607 Optics: Calibration Failed Restart the laser imager. See page 2-4.
If the error persists, call for service.
611 27611 Optics: Internal Hardware
Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
646 27646 Optics: Internal Hardware
Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
650 27650 Optics: Internal Hardware
Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
501 28501 Processor: Internal
Hardware Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
509 28509 Processor Warm-up Failure Restart the laser imager. See page 2-4.
If the error persists, call for service.
510 28510 Processor: Internal
Hardware Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
551 28551 Processor Heater Failure Restart the laser imager. See page 2-4.
If the error persists, call for service.
554 28554 Processor Over
Temperature
Restart the laser imager. See page 2-4.
If the error persists, call for service.
924 29924 Densitometer: Internal
Diagnostic Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
925 29925 Densitometer: Internal
Diagnostic Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
931 29931 Densitometer: Internal
Communications Failure
Restart the laser imager. See page 2-4.
If the error persists, call for service.
935 36935 Local Panel: No
Communications from MCS
Restart the laser imager. See page 2-4.
If the error persists, call for service.
Display
Screen
Web
Portal
Web Portal
Message
Action