Using and Maintaining the Imager
January 14, 2008 2G0733 2-11
Running a
Calibration Print
You may occasionally have to request that the Imager print a calibration
print to calibrate the Imager. You may be asked to do this if you call for
service. You should also run a calibration print if a "Not Calibrated" status
message appears on the Web Portal Home screen, or if a calibration error
code (2x-624, 2x-631, or 2x-632) appears on the display screen.
To request a calibration print from the local panel, press the calibration
button for the applicable film tray (see “Local Panel Layout”).
To request a calibration print from the Web Portal:
1. Access the Web Portal (see “Accessing the Web Portal”).
2. Click Calibrate, for the applicable film supply.
Calibration Failure Occasionally, the Imager may fail calibration and will notify the operator
with an error code on the display screen and a code and message on the Web
Portal. The most common cause is a film related problem. Depending on the
cause, the Imager may be able to continue operating, but would display the
"manual mode" icon.
Requesting a Density
Test
The Imager can print an internally generated density test print with a
SMPTE pattern. Density test prints are often used as a quality assurance tool
to verify the uniformity of films printed by the Imager.
To request a density test print:
1. Access the Web Portal (see “Accessing the Web Portal”).
2. Select Diagnostics>Test Print.
3. For "Select Test", use the drop-down menu to select Density Test.
4. Enter an optional label to be printed on the test film.