Troubleshooting
January 14, 2008 2G0733 3-17
Film Jam - Code
2x126
1. Remove the film trays (see “Inserting the Film Saver and Removing
the Film Tray”).
2. Remove any misplaced films from Area 1. This area can be accessed
through the film trays slots.
3. If Area 1 did not contain misplaced films, take the film trays to a dark
room and remove the film saver. Remove any misplaced films and
replace the film saver.
4. If misplaced films were not found in the tray, open and close the
Imager hood. Wait 2 minutes for misplaced films to emerge from the
Imager.
If misplaced films were never located and this error recurs on the next
film, continue with Step 5.
Film tray slots