Troubleshooting
January 14, 2008 2G0733 3-19
7. Remove any misplaced films seen in the pickup access areas.
8. Set the left panel in place and tighten the 2 screws.
9. Close the hood.
Film Jam - Code
26325
IMPORTANT: If this error occurred at the same time as 2x-126, follow the
instructions for “Film Jam - Code 2x126”.
1. If this error occurred alone, open and close the hood to clear the error.
2. If this error repeats, restart the Imager.
Film Jam - Codes
26326 or 26543
1. Open the hood.
2. Remove any misplaced films.
CAUTION:
Hot surface.
3. Undo the drum cover latches and lift the drum cover.
Hood
Drum cover latches
Drum cover