User's Guide
Table Of Contents
- 1 Overview
- 2 Using and Maintaining the Imager
- 3 Troubleshooting
- 4 Film Technical Information
- 5 Specifications
- Glossary
- Appendix A: Warranty and Limitation of Liability
- Important Notice to Purchaser
- End User License Agreement (EULA)
- Carestream Health, Inc. Software End User License Agreement
- End-User License Agreement for Microsoft Desktop Operating Systems
- Software Product License
- Warranty/Special Provisions for Austria, New Zealand or Papua New Guinea - Express Limited Warranty
- Warranty/Special Provisions for England Scotland, Wales and Ireland - Limited Warranty
- Warranty/Special Provisions for Canada - Limited Warranty
- GARANTIE ET DISPOSITIONS PARTICULIÈRES POUR LE CANADA GARANTIE LIMITÉE
- Warranty/Special Provisions for the United States of America and any Other Country - Limited Warranty
5-2 2G0733 January 14, 2008
Specifications
Other Location
Considerations
Additional factors that influence where the Imager is located include
proximity to patients, flammable materials, liquids, and other equipment.
CAUTION:
This equipment is not contained in a sealed cabinet. Do not use this
equipment in locations where it can come in contact with liquids,
including body fluids.
IMPORTANT:
• This equipment is intended to connect to other medical devices. Only
an authorized Service Provider of Carestream Health, Inc. Products or
Customer Qualified Service Personnel may install this equipment.
• Do not locate the Imager within 2 m (6.56 ft.) of where a cell phone
will be used, even if the cell phone is separated from the equipment
by a wall.
• Do not locate the Imager within 4 m (13.12 ft.) of a microwave oven.
Electromagnetic radiation from a microwave oven is only an issue if,
after the oven door is closed and latched, the seal does not maintain
an electromagnetic tight fit between the oven door and the oven main
housing. Determining if the seal has an electromagnetic tight fit
requires special detection equipment.
• Do not locate the Imager in the presence of flammable anesthetics,
oxygen, or nitrous oxide. The Imager does not have a gas-sealed
electronics enclosure and could ignite any flammable or explosive
gases present in the Imager’s environment.
• Do not locate the Imager in close contact with MRI devices due to
possible high magnetic fields near an MRI unit. The magnetic field in
the area where the Imager is installed must be less than 50 Gauss.
• Do not locate the Imager closer than 1.83 m (6 ft.) from a patient bed
or chair.
• Do not substitute or modify any part of the Imager without prior
written approval of Carestream Health, Inc.
1.83 m (6 ft.)