5-4 2G0733 January 14, 2008
Specifications
Power Requirements
The Imager power supply has an auto-sensing and auto-switching feature for
both voltage and frequency. The current draw is 10 A at 120 V AC, and the
power consumption is a maximum of 1275 W.
One of the following single-phase, 15 A power sources, with grounding,
must be provided within 2.5 m (8 ft.) of the Imager.
The wire must be insulation-rated for 600 V. A dedicated line is
recommended.
Network Requirements
The Imager receives digital images from medical imaging devices
(modalities) over a 10Base-T or 100Base-T ETHERNET Network. A single
CAT 5 UTP cable is provided in the accessories package for the Imager; 2
m (6.5 ft.).
Film
Prior to installation you must order and have on hand DRYVIEW Laser
Imaging Film. Contact your sales representative if you have questions.
Film Types The Imager accommodates the following film types:
KODAK DRYVIEW DVB Laser Imaging Film
KODAK DRYVIEW DVC Laser Imaging Film
KODAK DRYVIEW DVB+ Laser Imaging Film
KODAK DRYVIEW DVB+ Premium Laser Imaging Film
NOTE: Not all film types are available in every country.
Film Sizes The Imager accommodates the following film sizes:
20 x 25 cm (8 x 10 in.)
25 x 30 cm (10 x 12 in.)
28 x 35 cm (11 x 14 in.)
35 x 35 cm (14 x 14 in.)
35 x 43 cm (14 x 17 in.)
Film Storage Film must be stored at 5 - 25° C (41 - 77° F) and 85% RH or less.