User's Manual

Using and Maintaining the Imager
February 27, 2009 2G0733 2-15
3. Raise the hood.
4. Pull the front panel forward.
5. Remove the charcoal filter by lifting it up and pulling it forward.
6. Install a new charcoal filter.
7. Close the front panel.
8. Close the hood.
9. Install the film trays.
10. Remove the film savers.
11. Turn the Imager ON.
12. Wait for the Imager to complete the start-up routine.
13. Clear the preventive maintenance icon and condition code.
a. Press and hold the Menu Selection Button for 5 seconds. Release
the button and the Test Print menu displays.
b. Press and release the Menu Selection Button again to display the
Error Reset menu.
c. Check that the 20449 error is highlighted on the display screen,
and press the Enter button.
The icon and condition code are cleared, and the display screen
returns to normal operation.