User's Manual

Safety and Related Information
12 AA2205_en | 2012-07-27
Surge:
EN61000-4-5
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
lines:
EN61000-4-11
IEC 61000-4-11
<5 % U
τ
1
(>95 %
dip in U
τ) for 0.5
cycle
40 % U
τ (60 % dip
in U
τ) for 5 cycles
70 % U
τ (30 % dip
in U
τ) for 25 cycles
<5 % U
τ (>95 %
dip in U
τ) for 5 sec.
<5 % U
τ (>95 %
dip in U
τ) for 0.5
cycle
40 % U
τ (60 % dip
in U
τ) for 5 cycles
70 % U
τ (30 % dip
in U
τ) for 25 cycles
<5 % U
τ (>95 %
dip in U
τ) for 5 sec.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Imager requires continued operation
during power mains interruptions, it
is recommended that the Imager be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz) magnetic
field:
EN61000-4-8
IEC 61000-4-8
3 A/m 3 A/m Mains power quality should be that
of a typical commercial or hospital
environment.
1. Uτ is the a.c. mains voltage prior to application of the test level
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment—
Guidance
SHORT TERM CONFIDENTIAL