2010-11-04 2G8647 6-3
Environmental Requirements
Temperature Operating: 15° to 35°C (59° to 95°F)
Storage: -35° to 60°C (-31° to 140°F)
Relative Humidity Operating: 20% to 85% RH, noncondensing
Storage: 10% to 90% RH, noncondensing
Altitude 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
Floor Vibration For the Imager to operate with no image degradation, floor vibration
levels cannot exceed the following.
Floor Levelness The floor must be level within 1°.
G
2
/Hz Freq. Range Direction
1 x 10
-7
1-50 Hz Vertical or horizontal
6 x 10
-5
50-200 Hz Vertical or horizontal
5 x 10
-6
200-650 Hz Vertical or horizontal
1 x 10
-6
650-1000 Hz Vertical or horizontal