User's Manual
Table Of Contents
- 2G8647_001.pdf
- 2G8647_002.pdf
- 2 Using and Maintaining the Imager
- Operator Control of the Imager
- Turning Imager Power ON and OFF
- Power ON
- Normal Power OFF
- Emergency Power OFF
- Power Failures
- Power Save
- Using the Touch Screen
- Touch Screen Functions
- Touch Screen Help
- General Help
- Specific Screen Help
- Installation Road Map
- Physical Setup
- Configuration
- Imager Passcode
- Loading and Removing the Film Cartridges
- Drawers and Doors
- Opening Film Drawers and Main Door
- Opening the Film Registration Door
- Opening Air Intake Door
- Film Drawers and Main Door Manual Override Latch
- Filter Maintenance
- Cleaning the Electronics Filter
- Changing the PMM
- 2G8647_003.pdf
- 2G8647_004.pdf
- 4 Adding a Modality and Configuring Modality Image Quality
- Introduction
- Adding a Modality
- Configuring Modalities Automatically
- Configuring Modalities Manually
- Configuring Modalities that use PLUTs
- Configuring Modalities that use TFT Sets and Contrast Tables
- Fine Tuning the Image with TFTs Sets and Contrast Tables
- Performing a Contrast Test
- 2G8647_005.pdf
- 2G8647_006.pdf
- 6 Specifications
- Location
- Equipment Specifications
- Operating Space Requirements
- Other Location Considerations
- Environmental Requirements
- Temperature
- Relative Humidity
- Altitude
- Floor Vibration
- Floor Levelness
- Environmental Effects
- Power Requirements
- Network Requirements
- Film
- Film Types
- Film Sizes
- Film Storage and Throughput
- 2G8647_gls.pdf
Specifications
2010-11-04 2G8647 6-3
Environmental Requirements
Temperature • Operating: 15° to 35°C (59° to 95°F)
• Storage: -35° to 60°C (-31° to 140°F)
Relative Humidity • Operating: 20% to 85% RH, noncondensing
• Storage: 10% to 90% RH, noncondensing
Altitude • 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
Floor Vibration For the Imager to operate with no image degradation, floor vibration
levels cannot exceed the following.
Floor Levelness The floor must be level within 1°.
G
2
/Hz Freq. Range Direction
1 x 10
-7
1-50 Hz Vertical or horizontal
6 x 10
-5
50-200 Hz Vertical or horizontal
5 x 10
-6
200-650 Hz Vertical or horizontal
1 x 10
-6
650-1000 Hz Vertical or horizontal