User Manual

2-4 7H8166
Hardware and Operation
Attaching Accessories The use of equipment and/or hardware that does not comply with the
equivalent product safety and EMC requirements of this product may lead to a
reduced level of safety and/or EMC performance of the resulting system.
Consideration relating to the choice of accessory equipment used with this
product shall include:
Use of the accessory in the patient’s vicinity.
Evidence that the safety certification of the accessory has been
performed in accordance with applicable coordinated harmonized
product safety standards per IEC 60601-1-1.
Evidence that applicable emission certification of the accessory has been
performed.
Turning the System
On and Off
To turn the DRX-1 System On:
1. Press the ON switch on the UPS.
2. Press the ON switch on the computer and monitor.
3. When the software initializes, select the DRX-1 icon on the monitor.
To turn the DRX-1 System Off:
1. Select the Quick Menu in the lower left corner of the monitor.
2. Select Shut Down.
3. Turn Off the monitor.
4. Turn Off the UPS.
You can remain in hold-on-prep for up to 15 seconds with the CARESTREAM
DX-1 System Detector and the detector will function properly. In the event of
an aborted exposure, the detector can acquire a subsequent image in four
seconds.
Carestream
Carestream SystemDRX1
UPS