User Manual
Table Of Contents
- 1 Safety and Regulatory Information
- Document Conventions
- Intended Use and Indications for Use
- Safety and Related Information
- DRX-1 System Product Information
- Disposal Information
- Operating Environment
- 2 Hardware and Operation
- Overview
- CARESTREAM DRX-1 System
- Cautions
- Installing the Hardware
- CARESTREAM DRX-1 System Battery
- Labeling the Detector
- DRX-1 Detector LED
- Positioning the Detector in the Bucky
- Range of Operation
- Wireless Operation
- Tether Operation
- Cleaning the Hardware
- With Each Occurrence of Patient Contact
- System Maintenance
1-2 7H8166
Safety and Regulatory Information
Safety and Related Information
Medical Equipment
Classification
CARESTREAM DRX-1 System Detector Medical Electrical
Equipment Classification
Type of protection against electrical shock: Internally powered equipment. Class I
Equipment.
Degree of protection against electrical
shock:
Type B Applied Part.
Degree of protection against ingress of wa-
ter:
Ordinary protection.
Mode of operation: Continuous operation.
Flammable anesthetics: Not suitable for use in the presence of flam-
mable anesthetics or a mixture of flamma-
ble anesthetics with air or oxygen or
nitrous oxide.
CARESTREAM DRX-1 System Tether Interface Medical Electrical
Equipment Classification
Type of protection against electrical shock: Class I Equipment.
Degree of protection against electrical
shock:
Type B.
Degree of protection against ingress of wa-
ter:
Ordinary protection.
Mode of operation: Continuous operation.
Flammable anesthetics: Not suitable for use in the presence of flam-
mable anesthetics or a mixture of flamma-
ble anesthetics with air or oxygen or
nitrous oxide.