User's Manual

10
Communications Equipment
Portable and mobile radio frequency (RF) communications equipment
can affect medical electrical equipment EMC performance.
The wireless version of the Kodak 1500 intraoral camera operates with a
802.11g protocol in a 2.4GHz frequency band. The radio output power
is 20 mW (nominal).
Accessories
The use of cables, adapters, or accessories other than those specified
with the exception of those sold by the manufacturer of the equipment as
replacement parts for internal components, may result in increased
emissions or decreased immunity of the medical equipment.
Changes or modifications not expressly approved by the party responsible
for compliance could void the user's authority to operate the equipment
Other Equipment
The Kodak 1500 intraoral camera should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is necessary,
the Kodak 1500 intraoral camera should be observed to verify normal
operation in the configuration in which it will be used.
Condition Classification
Type of protection against electrical shock: Class II Equipment
Degree of protection against electrical shock: Type BF applied part
Degree of protection against ingress of water: Ordinary Protection
Flammable anesthetics:
Not suitable for use in the
presence of flammable
anesthetics or a mixture of
flammable anesthetics with air or
oxygen or nitrous oxide.
Mode of operation Continuous operation