5 Trigger the x-ray with the remote control of the x-ray generator. The image appears in the preview
screen of your device.
The light on the remote control button blinks blue indicating the image
transmission
6 When the acquisition ends, the light on the remote control button becomes solid green. At this stage,
while the count down continues, you can continue acquiring other images.
If the count down ends and you wish to acquire other images then redo step 5.
7 Check the image quality. If not satisfactory, redo the x-ray. If satisfactory, remove the generator
tube head.
8 Remove the RVG sensor from the patient’s mouth. Remove the hygienic sensor protection.
9 Tap on the acquired image to access the contrast tools. The acquired images is saved automatically.
Important: DO NOT pull the sensor by its cable when you remove
the hygienic protection.