User Manual

Table Of Contents
KODAK RVG 6500 System_User & Installation Guide (SM728)_Ed01 61
Information Messages
An information “I” error code with a message is displayed on the popup on your computer screen.
The following table lists the information messages and the action to take.
Table 3 Information Messages
Information
Error Code
Information Message Action
I 1
Communication problem with the wireless RVG
sensor. Turn off then on your sensor or restart
your computer. If the problem persists, contact
your representative.
Turn Off then ON the RVG sensor.
Or reboot your computer.
If the problem persists, contact your
representative.
I 2
Wireless sensor error. Turn off then on your
sensor and restart the acquisition software. If
the problem persists, contact your
representative.
1. Turn Off then ON the RVG sensor.
2. Click to access the RVG
Acquisition interface
.
If the problem persists, contact your
representative.
I 3
Error during image acquisition. Turn off then on
your wireless sensor and restart the acquisition
software. If the problem persists, contact your
representative.
1. Turn Off then ON the RVG sensor.
2. Click to access the RVG
Acquisition interface
.
If the problem persists, contact your
representative.
I 4
Memory allocation problem. Turn off then on your
wireless sensor. If the problem persists, contact
your representative.
Turn Off then ON the RVG sensor.
If the problem persists, contact your
representative.
I 5
Wireless sensor has storage/transmission
memory problem. Contact your representative.
Contact your representative.
I 6
Incorrect IPS calibration. Contact your
representative
Contact your representative.
I 7
IPS Batteries have run down. Change the
batteries.
Change the batteries of the IPS aiming
ring.
I 9
Writing transmission error. Turn off then on your
wireless sensor and restart the process. If the
problem persists, contact your representative.
1. Turn Off then ON the RVG sensor.
2. Redo the process
If the problem persists, contact your
representative.
I 10
Reading transmission error. Turn off then on your
wireless sensor and restart the process. If the
problem persists, contact your representative.
1. Turn Off then ON the RVG sensor.
2. Redo the process
If the problem persists, contact your
representative.