Use Instructions

20
Instructions for Use | Anti-Decubitus Air Alternating Pressure Mattress Replacement System
AVA Activity Analytics
Mute
User may press this button to turn o󰀨 the audible indicator. After mute,
the audible indicator will not re-start again and only have LED ashing to
visually notify the caregivers until the problem is solved.
Panel Unlock
When there is no operation after 2 minutes, system will automatically
lock the control panel. The white LED will light to indicate the panel is
locked. All buttons will be locked except the Max. Inflation and Mute
button which allow user to mute audible alert. Simply press and hold this
button for 3 seconds to release it from locking.
Failure Indicator LED
This yellow LED is to indicate that the mattress has abnormal pressure
including insu󰀩cient pressure, excessive pressure and service required
including bottoming out,power issue,rator valve issue or sensor
communication issue. There will be both audible beeping and visual LED
indicators. Audible indicator will turn o󰀨 after pressing the Mute button
and it will not start beeping again to avoid acoustic pollution, but the LED
will stay ashing until the problem has been solved.
WiFi Indicator
There are three kinds of LED colors and four kinds of conditions of Wi-
Fi indicator. Please refer another IFU to learn how to set the Wi- Fi
Connection.
(1) Amber. There is no Wi-Fi connection setting via mobile APP.
(2) Blinking blue. Power unit is trying to connect to Wi-Fi router
(3) Blinking white. Power unit is trying to connect to server.
(4) Static white. Power unit is connected.
(5)
LED is o󰀨: Turn o󰀨 the Wi-Fi function and become standalone version.
IMPORTANT
If no action has been done in 30 minutes, the device will turn into
standalone version.
IMPORTANT
Reset power unit for Wi-Fi settings via press power button for 10
seconds.
Digital Number display - Envelopment rate
Show Envelopment Rate in the middle digital number display of the
panel. The number will be changed via revolving the knob.
• IMPORTANT :
Strongly suggest that the envelopment rate must higher than 50 to
keep the best theraphy and the accurate sensor detection.
Digital Number display - Cycle Time
Show Cycle time or countdown time in the left side digital number display
of the panel.
5
Features