Use Instructions

22
Instructions for Use | Anti-Decubitus Air Alternating Pressure Mattress Replacement System
AVA Activity Analytics
Specication ID Label (3)
This label includes all the information for medical device and safety
requirements.
Air Filter (4)
We recommend inspecting and cleaning this filter monthly or more often,
depending on the environment the unit is being used. Failure to keep the
filter clean will result in shorter life span of the unit and/or unit failure. When
replacing, be sure to use Carilex standard lter to optimize the performance of
the power unit.
6.1 System Installation
(1) Remove the existing mattress.
(2) Put the mattress on the structure of the bedframe with the logo at
the foot of the mattress. Fasten the mattress with the use of the
straps ensuring that the functions and movements of the bedframe
are not limited before proceeding to the next step.
(3) Hang the power unit at the foot of the bedframe with aid of the
hanger hooks.
(4) Push the quick coupling of the mattress rmly into the air outlet of
the power unit.
(5) Push in the transfer cable of the mattress firmly into the rapid
connector of the power unit.
(6) Insert the power cord set into a wall outlet. Ensure that the cable is
not in the way of the operators and the movement of the bedframe.
(7) Check the air cell holder of mattress is at without any fold before
initiating and inating the mattress.
(8) Check 26pcs rubbers of sensors’ pad are protruding.
(9) The power button is bright in static white.
(10) Push the power button to initiate the system.
• IMPORTANT
The initial process will take longer than35 minutes.It takes 35 to 45
minutes depends on mattress size.
6.2 Operation
(1) Initiate the system by pressing the power button and wait for the
pump beginning to inate the mattress air cell. AVA Activity Analytics
is equipped with Carilex’s IDS technology,and the system will
automatically provide the optimal envelopment rate to patient. Once
the patient is on the mattress,the system will automatically adjust
and customize the pressure setting that is most suitable for the
individual patient load and position.
(2) Ensure to select the appropriate therapy mode and cycle time,
according to the physician’s decision.
(3) IMPORTANT:
Tucking in the coverlet too tensely signicantly reduces the
mattress' e󰀨ectiveness.
6
System Installation,Operation,Transfer and Transport of Patient