Instructions for Use | Anti-Decubitus Air Alternating Pressure Mattress Replacement System
AVA Activity Analytics
Problem Control Procedure Possible Solution
4.2) Check if the mattress is
insu󰀩cient pressure.
4.2) Check if there is any air loss
from air cell.
4.3) Check if the mattress is
excessive pressure.
4.3) Check if the mattress is too
hard.
4.4) Check if the power cord is loose. 4.4) Check the power cord is rmly
connect with power unit and wall
socket.
4.5) The alternating valve is not
working correctly.
4.5) Contact the authorized
distributor for technical service.
5) Wi-Fi disconnect fail 5.1) See Wi-Fi indicator LED is
amber.
5.1) Check you have set the Wi-Fi
connection between power unit, Wi-
Fi router and server via mobile APP.
(Please refer another IFU)
5.2) See Wi-Fi indicator LED is
blinking in blue.
5.2) Check the Wi-Fi router is on
and work well.
5.3) See Wi-Fi indicator LED is
blinking in white.
5.3) Check the server is on and
intranet connection work well.
6.) Initial calibration fail 6.1) Check patient is not on the
mattress before initial calibration
ready.
6.1) Remove patient carefully until
initial calibration ready or press “Max
inate” to ignore the fail.
6.2) Check the mattress is placed at
on the bedframe and all air cells are
not extrusion or bending.
6.2) Try to replace the mattress
again for a at condition,
6.3) Check the air cell holder of
mattress is not folded.
6.3) Use hand to check the air cell
holder of the mattress.
7.) Sensor pairing fail 7.1) Check the transfer cable is
connected before inserting the power
cord set into a wall outlet.
7.1) Reconnect the transfer cable
with rapid connector of power unit
7.2) Still see the failure indicator,
power LED is blinking in white,and
control panel shows the number via
digital number display.
*Number(N) could be 01~27
*Refer below sensor pad drawing
7.2) Remove power cord
7.3) Check if the No.N sensor is
disconnected form the sensor board
cable of mattress sensor’s pad.
7.3) Push the cable into the sensor
board connector rmly and you
should feel a ”click”.
7.4) Reconnect the power cord
Troubleshooting
10