User Manual

PAGE 1
Issue Rev A 1208
Quincy Lab Inc. 1925 N. Leamington Ave. Chicago, Illinois 60639 1-800-482-HEAT (4328)
Copyright Quincy Lab Inc. 2008
Model Series 140 & 180
General Purpose Incubators
OPERATING MANUAL
Quincy Lab, Inc.
H
EA
T/C
YC
LE
POWER
Model 12-140 Incubator
Interior Dimensions
INCHES W x H x D
(CM) W x H x D
Exterior Dimensions
INCHES W x H x D
(CM) W x H x D
12x10x10 18x16x12
18x16x1212x10x10
13x15x11
19x21x13 13x15x11
19x21x11
Weight (lbs)
19 lbs 33 lbs 19 lbs
33 lbs
Standard Electrical
31x25x25 46x41x30 46x41x3031x25x25
33x38x28
48x53x33
33x38x28
48x53x33
SPECIFICATIONS
MODEL
10-140
MODEL
12-140
MODEL
10-180
MODEL
12-180
Temperature Range
Ambient + 2
o
C to 62
o
C
Common Unit Construction
Exterior: Powder-Coated Steel Interior: Aluminum
Insulation: Fiberglass Door: 140: Acrylic, 180: Steel Insulated
Thermo-control: Bi-Metal Heater: Resistive-Tubular Incoloy
* Standard models voltage only, optional 230 voltage available. Check label on back of unit.
VOLTS / WATTS
115 / 120*
115 / 235*
115 / 270* 115 / 385*
HEAT/CYCLE
POW
ER
M
o
d
e
l 10
-1
80
In
c
u
b
a
to
r
Ambient + 3
o
C to 94
o
C
Cubic Foot Capacity
.7 ft
3
2.0 ft
3
.7 ft
3
2.0 ft
3
Model 10-180
Model 12-140
Common Unit Specifications
Operating Environment: Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,
Pollution Degree 2, ambient temperature 10
o
C/50
o
F to 35
o
C/95
o
F,
80% RH maximum.
Storage Temperature: -10
o
C/14
o
F to 70
o
C/158
o
F, 70% RH maximum.
Approvals: Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)
Compliance: UL Standard 61010-1, IEC 61010-1, 2nd Edition.