TEST_30 - CPD Registration
6. The MS Database is read and the following data is displayed:
Phone number: <decimal number>
MCC: <hexadecimal number>
MNC: <hexadecimal number>
Serial number: <decimal number>, <decimal number>, <
hexadecimal number>
Power Class: <class I, II, or III>
Transmission: <continuous or discontinuous>
Slot Class: <non-slotted or slotted>
Slot Index: <decimal number>
Dual Mode <dual mode or CDMA only>
NOTE
Regarding the serial number data shown above, the numbers are as
follows:
First Decimal Number: The first three digits represent the
manufacturer’s code. The next eight digits represent a combination of
the reserve bits and the unit serial number.
Second Decimal Number: The first three digits represent the
manufacturer’s code. The next two digits represent the reserve bits. The
next six digits represent the unit serial number.
Hexadecimal Number: A hexadecimal representation of the above two
numbers.
Refer to the Agilent 8924C Reference Guide for information about the
Sector A Power, RF Channel, Register, MS ID, and MS Database fields.
NOTE
A TEST_30 - CPD Registration or a TEST_31 - CPD Origination must
be run at least once before any digital tests can be run. The Test Set
remembers the registration data unless the power is cycled on the Test
Set. If power is cycled on the Test Set, the TEST_30 - CPD Registration
or the TEST_31 - CPD Origination must be run again before running
other digital tests.
Pass/Fail Limits Used
None