Specifications

7890 Series GC Site Preparation
GC, GC/MS, and ALS Site Preparation Guide 13
Dimensions and Weight
Select the laboratory bench space before the system arrives. Make sure the
area is clean, clear, and level. Pay special attention to the total height
requirements. Avoid bench space with overhanging shelves. See Table 2.
The instrument needs space for proper convection of heat and ventilation.
Allow at least 25 cm (10 in) clearance between back of the instrument
and wall to dissipate hot air and allow for routine maintenance.
Table 2 Required instrument height, width, depth, and weight
Product Height Width Depth Weight
GC
7890 Series GCs
With third detector
50 cm (19.2 in)
50 cm (19.2 in)
59 cm (23 in)
68 cm (27 in)
54 cm (21 in)
54 cm (21 in)
50 kg (112 lb)
57 kg (125.4 lb)
GC operational oven access Requires 30 cm (12 in) open space above GC
Ion Trap MS
220 Ion Trap MS 49 cm (19 in) 25 cm (10 in) 64 cm (25 in) 23 kg (51 lb)
Foreline pump (with oil mist
eliminator)
38 cm (15 in) 21 cm (8 in) 31 cm (12 in) 11 kg (24 lb)
GC/Ion Trap MS operational and maintenance access Requires 76 cm (30 in) above the instrument and 22 cm (9 in)
to the right
240 Ion Trap MS 49 cm (19 in) 38 cm (15 in) 66 cm (26 in) 42 kg (93 lb)
Foreline pump (with oil mist
eliminator), standard
Oil-free
46 cm (18 in)
19 cm (7.5 in)
21 cm (8 in)
32 cm (13 in)
43 cm (17 in)
28 cm (11 in)
22 kg (49 lb)
16 kg (35.2 lb)
GC/Ion Trap MS operational and maintenance access Requires 76 cm (30 in) above the instrument and 22 cm (9 in)
to the right
MSD
5975 Series MSD
Diffusion pump 41 cm (16 in) 30 cm (12 in) 54 cm (22 in) 39 kg (85 lb)
Standard turbo pump 41 cm (16 in) 30 cm (12 in) 54 cm (22 in) 39 kg (85 lb)