Specifications

7890 Series GC Site Preparation
30 GC, GC/MS, and ALS Site Preparation Guide
MS
7000 Triple Quad MS Operation 15 to 35 °C
(59 to 95 °F)
40 to 80% 5,000 m
Storage –20 to 70 °C
(–4 to 158 °F)
0 to 95%
7200 Q-TOF MS Operation 15 to 35 °C
(59 to 95 °F)
20 to 80% 5,000 m
Storage –20 to 70 °C
(–4 to 158 °F)
0 to 95%
* Operation requires constant temperature (variations < 2
o
C/hour)
Operation requires constant temperature (variations < 2
o
C/hour)
An altitude of 3,700 meters (12,000 feet) is supported if the ambient temperature is less than 30 °C
Table 6 Environmental conditions for operation and storage (continued)
Product Conditions Operating temp
range
Operating
humidity range
Maximum
altitude