User's Manual

AG501 cs5cal (PD-02.05)
Page 6 of 11
3. Select the sensors to calibrate and start the calibration
Fig.14: Select the magazines that contain sensors for calibration and
check if all sensors are in their right position
Please note: As the calibration program calculates the relation between the four sensors in a
magazine, it is recommended to completely fill the magazine so it is proposed to calibrate 4, 8, 12 or
16 sensors (6, 12, 18 or 24 sensors for 24 channel systems respectively) in each case four (or six)
sensors in one magazine. Please make sure that the sensors are placed in the right order. Then click
OK to start the data recording.