E-box firmware update (Z2-FPGA)
Page 1 of 1
AG501 E-Box
This description only applies to AG501-systems with the Z2-FPGA implemented in the E-Box.
AG501-systems with an installed Microblaze obtain firmware updates via the controlserver.
If in doubt, which update procedure applies to your system, please contact Carstens
Medizinelektronik.
Changing the SD-card of the Z2-FPGA
The firmware of the Z2-FPGA is stored on an SD-card mounted to the FPGA.
So to update the firmware the SD-card has to be updated or exchanged for an SD-card with the
desired firmware.
1. Turn off the E-Box and disconnect the power connector.
2. Screw off the cover of the E-Box (Figure 1 A).
3. Replace SD-card in slot with an SD-card containing the new firmware (Figure 2).
Make sure the SD-card sits tightly.
4. Close E-Box and tighten screws.
5. Reconnect E-Box and test system.
Figure 1: Take of the E-Box-cover Figure 2: The SD-card is mounted to the Z2-FPGA
A
SD
-