Digital Electronic Coating Thickness Gauge DPM-816 PRO USER MANUAL EN-01.01.
Contents - Device controls 3 - Introduction 5 - Purpose 6 - Application 6 - Device care 7 - Cleaning the device 8 - Storage conditions 8 - Display 8 - Sound signal 8 - Sensor 9 - Included in the set 9 - Getting started 9 - Installing the battery 10 - Low battery indicator 10 - Turning on/off the device 10 - Turning on/off the backlight 11 - Operating modes 12 - Units of measurement 13 User Manual DPM-816 PRO Page 2
- Taking the measurement 13 - Calibration 15 - Zero deviation calibration (one-point calibration) 16 - Removing calibrations 17 - Two-point calibration 18 - Six-point calibration 19 - Checking the car body 22 - Specifications 23 - Possible malfunctions and suggested remedies 25 - Warranty 26 Device controls "MODE" button - Switches the operating mode - Increases the thickness during calibration - Switches on two-point calibration - Switches on six-point calibration of ferrous metals "POW
- Switches the unit of measurement - Decreases the thickness during calibration - Resets zero deviation and two-point calibrations - Switches on six-point calibration of non-ferrous metals User Manual DPM-816 PRO Page 4
Introduction Dear owner of the digital coating thickness gauge (referred to as "the device" from this point), we congratulate you with the purchase of the device of Russian origin. We are confident that the device will be useful to you, and last a long time. For correct and optimal use of the device, read this manual carefully.
- Small and convenient size - Automatically turns off power and backlight - 5 modes for various measurement conditions - Independent calibration using one, two or six points of reference, saved to non-volatile memory of the device - Simple, intuitive controls with 3 buttons - Distinct sound signal for each measurement method and button press - Complies with industry standards - 3 year factory warranty Purpose The device is designed to measure coating thickness: Magnetic induction method - thickness of vari
Diagnostics of the car body to assess its condition. For example: when buying a car, before repairing dents without paint (PDR), or scratch removal. Device care Please keep in mind that the device is a high precision measuring instrument. Avoid dropping the device and subjecting it to mechanical impact. The device is protected against moisture, but is not waterproof. It must not be used underwater or in heavy rain.
Cleaning the device To remove dust from the device case or the sensor, use a soft, dry cloth. Do not use cleaning agents containing organic solvents. Storage conditions Store the device in a warm and dry place away from direct sunlight. Do not leave the battery inside the device during long term storage. Do not leave the device near strong sources of magnetic fields, for example, near magnets, power supplies or electric motors.
substrate under the coating when measuring in automatic mode, without having to look at the display. Sensor To increase the accuracy of measurements and to enable measurements on convex and concave surfaces, the device uses a sensor constructed to stabilize the pressure on its sensitive tip against the surface, as well as a spherical form for its sensitive tip. The sensitive part of the sensor is made of high-strength heattreated steel, and the area of contact with the measured surface is less than 1 mm².
Before first use, or in case the device has not been used for an extended period, as well as to increase the measurement accuracy, you need to calibrate the device using one or two points of reference. Installing the battery Open the battery cover, shifting the cover with your thumb in the direction of the arrow. Insert the battery, taking note of the polarity. Close the battery cover until it clicks.
The device will begin to automatically test and calibrate the sensor. After 2-3 seconds, the display will show dashes, indicating that the device is ready to take measurements. IMPORTANT! When turning on the device, keep it away from metal objects and strong sources of magnetic fields! Turning off the device To turn off the device from any state, press and hold the button for 2 seconds. Display backlight The device has an integrated backlight for the display.
Operating modes The device has 5 operating modes: A, M, P, F(Fe), N(nFe). To switch between these modes, shortly press the button, which will switch modes in a loop. The current mode will be shown on the display. A - Auto: Quick measurement with automatic selection of measurement method. Dynamic resolution. Great for checking the car body. M - Manufacturing: Detailed and precise measurements with high resolution across the entire range. Automatic selection of measurement method.
N(nFe) - Eddy current method: Quick, continuous measurement (4 measurements per second) of coating thickness on an aluminum alloy substrate, using eddy current. For accurate measurements, it is recommended to complete the two-point calibration before measuring. Units of measurement For user convenience, the device can display the results of measurement in several units of measurement: millimeters, mils, and micrometers.
Taking measurements in A, M, P modes (automatic modes): Firmly place the device sensor flat against the surface to be measured. Do not move the device during measurement (until a sound signal is emitted), otherwise the measurement may be inaccurate. Once the device detects that the sensor is on the surface in a fixed position, you will hear a sound signal indicating the completion of measurement. Move the device away from the measured surface.
These modes are useful when it is difficult to hold the device in a fixed position. For example, when measuring on pipes, convex and concave surfaces. Calibration For accurate thickness measurements, the device needs to be calibrated. The device retains the calibration information and in most cases you don't need to calibrate it again before every use. Be sure to check and complete the calibration the first time you use the device, or if you have not used it for a long time.
Keep a clean environment during calibration, as the slightest bit of dust stuck between plastic and metal base plates may distort the result by 5-10 µm. Zero deviation calibration (one-point calibration) When to calibrate: 1. When measuring on a metal substrate without a coating, the device will show a result greater or less than zero (depending on the properties of the metal substrate). After calibration, the device will display zero. 2. In cases where you need to measure deviation from a sample.
Verify the result by repeating the measurement. Removing calibrations This function will completely remove zero deviation and two-point calibrations for all modes. This function will restore the device's functionality in case of incorrect calibration, but for accurate measurements it needs to be calibrated again. To remove calibrations, when the display shows dashes , press and hold the button for 3 seconds. The display will show "-CL-" and a sound signal will be emitted.
Two-point calibration When to calibrate: - Before first use; - If the device has not been used for an extended period; - Before each measurement for maximum accuracy. Two-point calibration procedure: To enter calibration mode, press and hold the button for 2 seconds until the display shows "F-0". This indicates the zero point calibration. Take a measurement on a metal substrate without a coating, for which the device will be calibrated. If necessary, repeat the measurement.
result is being displayed, shortly press the or buttons to adjust the displayed thickness of the measured coating, and shortly press the button to save the calibration. The display will show "8888", indicating that the calibration has been successfully saved. Verify that the calibration is correct by repeating the measurement of the metal substrate with and without the sample coating. If unsuccessful, repeat the calibration.
F-4 (point 4) - Plate with a thickness of 1900-2500µm (75-98 mils). F-5 (point 5) - Plate with a thickness of 2600-3500µm (102-137 mils). Entering the calibration mode: Turn off the device, then press and hold the button to enter calibration mode for a ferrous metal substrate (magnetic induction method) or the button to enter calibration mode for a nonferrous metal substrate (eddy current method), until the display shows "FCAL" or "ACAL" respectively.
The display will show "F-1" - the next point of calibration, and the device will show the recommended thickness of calibration plate for the current point of calibration. Take a measurement of the metal base together with the calibration plate. If necessary, repeat the measurement. Use the and buttons to calibrate the displayed thickness of the measured calibration plate, and shortly press the button to proceed to the next point. Repeat the procedure for each point of calibration.
Please note, if the calibration is done incorrectly on one of the metal base types, the device will be unable to automatically detect the type of substrate material and choose the suitable measurement method. The new device's six-point calibration is identical to the factory calibration. The device is calibrated at the factory using steel (magnetic induction method of measurement) and aluminum (eddy current measurement method) substrates.
Specifications Measuring range (normalized) - mm - µm - mil 0-3 0-3000 0-118 Maximum measurable thickness - mm - µm - mil 3.5 3500 137 Resolution (A mode) - mm (dynamic) - µm (dynamic) - mil (dynamic) 0.001-0.01 1-10 0.01-0.5 Maximum resolution (M, P, F, N modes) - mm (fixed) - µm (fixed) - mil (dynamic) 0.001 1 0.01-0.
mils – ±(3%+0.5) Operating conditions - ambient temperature, ºС - ambient temperature, ºF - relative humidity, % not exceeding -25..+40 -13..+104 80 Storage and transportation conditions - ambient temperature, ºС - ambient temperature, ºF - relative humidity, % not exceeding -25..+55 -13..+131 80 Supply voltage (battery type AAA), V 1.5 Current consumption, mA, not exceeding - in sleep mode (off state) - in measurement mode 0.5 60 Overall dimensions, - mm - in 104х43х29 4.1x1.7x1.
Possible malfunctions and suggested remedies Malfunction Probable cause Remedy The device does not turn on, or turns on and immediately turns off. Incorrect installation of the battery, reversed polarity. The battery is discharged, or its service lifespan has ended. Check the battery polarity. If necessary, replace the battery. The device turns off when the The battery is low, or its backlight is turned on, or the low service lifespan has ended. battery indicator is displayed.
When measuring on certain substrates, the thickness reading wrong. The current calibration preset is unsuitable for the current substrate. Complete a two-point calibration. When the device is placed against the measured surface, the measurement does not occur. Incorrect device positioning relative to the measured surface. The device must be firmly pressed against the measured surface and remain fixed until the sound signal is emitted.
The warranty period for "CARSYS DPM-816 Pro" is 3 years from the date of purchase of the device by the buyer. All the details are at https://car-sys.com/en/ The device is designed and manufactured by: CARSYS, Vladimir Chuvakin, Russia, 644007, Omsk, 80 Frunze st.