User Manual DPM-816 PRO Page 20
F-4 (point 4) - Plate with a thickness of 1900-2500µm (75-98 mils).
F-5 (point 5) - Plate with a thickness of 2600-3500µm (102-137 mils).
Entering the calibration mode:
Turn off the device, then press and hold the button to enter
calibration mode for a ferrous metal substrate (magnetic induction
method) or the button to enter calibration mode for a non-
ferrous metal substrate (eddy current method), until the display
shows "FCAL" or "ACAL" respectively.
Wait until the display shows "0", indicating zero point calibration.
From this moment, calibration can begin.
Calibration process:
The first point of calibration is "0". Take a measurement of the
metal base without a coating, for which the device will be
calibrated. If necessary, repeat the measurement.
Shortly press the button to save the result and proceed to
the next point of calibration.