Cascade TEK Division of Cascade Technical Sciences, Inc.
This instruction manual applies to Models TFO-1, TFO-3, TFO-5, TFO-10, TFO-28, TVO-1, TVO-2, and TVO-5 TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 WARNING SECTION 3.0 POWER SUPPLY SECTION 4.0 INSTALLATION SECTION 5.0 CONTROLS SECTION 6.0 OPERATION SECTION 7.0 PREVENTATIVE MAINTENANCE SECTION 8.0 TROUBLE SHOOTING SECTION 9.0 PART LISTS SECTION 10.
AN INVITATION Please take a few minutes to peruse this instruction manual before using your new oven. We want you to be happy with the equipment, and use it safely. Please be sure that each person who uses the oven reads the instructions. 1.0 RECEIVING AND INSPECTION The carrier, when accepting shipment, also accepts responsibility for safe delivery, and is liable for loss or damage claims. On delivery, you must inspect for visible exterior damage.
Note: 240-volt ovens may be operated on 208-volts. Heat-up will be slightly slower. GROUND YOUR OVEN Our 120-volt ovens are fitted with a power cord having a plug with grounding blade. Be sure the plug is used with a matching receptacle and that the grounding blade is used. DO NOT cut off the grounding blade or plug into a non-grounded receptacle. This would leave the oven's cabinet ungrounded in violation of electrical codes and common sense.
Vacuum Gauge: Measures the vacuum level in the oven form 0 – 30” of vacuum. 6.0 HOW TO RUN THE OVEN The following instructions are for single set point operations only. The Watlow 981 controller has 4 automatic settable profiles with six steps within each profile. For programming and set-up instructions turn to Chapter 7 (page 7.1) in the Watlow manual supplied with your oven.
8.0 TROUBLE-SHOOTING No Heat: 1. The branch circuits serving the oven are dead. 2. Oven cord plug is not in its receptacle. 3. Oven power switch is open (off). 4. The temperature controller is set too low. 5. The overtemperature safety is set too low. It must be higher than the controller. 6. The controller's Thermocouple temperature sensor is open circuited. 7. The solid state relay is open circuited. 8. Line contactors have failed open circuited. (Model TFO-28) 9.
9.0 PARTS LIST FORCED AIR OVENS PART NO. DESCRIPTION TFO-1 1800516 9570762 9600565 100056 4880512 600002 5120689 200129 DESCRIPTION TFO-3 Power Cord Element Assy. Blower Motor Assy. Blower Wheel 6 x 2 Blower Motor Gasket 10 ft Shelf Adjustable Feet TFO-5 101990 9570665 890081 9600565 100056 4880512 600002 5120691 200129 PART NO. 1800516 9570537 9600565 100056 4880512 600002 5120690 200129 Power Cord Element Assy. Blower Motor Assy.
VACUUM OVENS TVO-1 9570739 5620506 5620519 Element Assy., top / bottom Shelf - Top 2 Shelf - Bottom 1 TVO-2 9570742 5620518 5620520 Element Assy. Shelf - Top 2 Shelf - Bottom 1 TVO-5 9570765 5620523 5220580 9570764 Element Assy., top / bottom Shelf Shelf Sliders Element Assy., side PARTS FOR ALL VACUUM OVENS 9530535 9530536 100031 1800516 X1000771 103351 200020 8000542 100026 101827 1750540 102162 200129 Needle Valve Kit .250 Needle Valve Kit .
North America Cascade TEK agrees to correct for the original user of this product, either by repair or at our election, by replacement, any defect in material or workmanship which develop within twelve (12) months after delivery of this product to the original user. In the event of replacement, replacement unit will be warranted for the original twelve (12) month period or ninety (90) days, whichever is longer. If this product should require service, contact Cascade TEK, Hillsboro, OR at (503) 648-1818.
TFO-1, TFO-3
WIRING DIAGRAM TFO-5
WIRING DIAGRAM TFO-10
WIRING DIAGRAM TFO-28