The trigger settings are available for R&S SMx and R&S AMU instruments only.
EXTernal<ch>
The numeric suffix to EXTernal<ch> distinguishes between the external trigger via the
TRIGGER 1 (suffix 1) and TRIGGER 2 (suffix 2) connector.
[:SOURce<hw>]:BB:GPS:TRIGger:ARM:EXECute............................................................132
[:SOURce<hw>]:BB:GPS:TRIGger:EXECute....................................................................132
[:SOURce<hw>]:BB:GPS:TRIGger:EXTernal:SYNChronize:OUTPut...................................133
[:SOURce<hw>]:BB:GPS:TRIGger:OBASeband:DELay.....................................................133
[:SOURce<hw>]:BB:GPS:TRIGger:OBASeband:INHibit.....................................................134
[:SOURce<hw>]:BB:GPS:TRIGger:RMODe......................................................................134
[:SOURce<hw>]:BB:GPS:TRIGger:SLENgth.....................................................................135
[:SOURce<hw>]:BB:GPS:TRIGger:SLUNit.......................................................................135
[:SOURce<hw>]:BB:GPS:TRIGger:SOURce.....................................................................135
[:SOURce<hw>]:BB:GPS:TRIGger[:EXTernal<ch>]:DELay.................................................136
[:SOURce<hw>]:BB:GPS:TRIGger[:EXTernal<ch>]:INHibit.................................................136
[:SOURce<hw>]:BB:GPS[:TRIGger]:SEQuence................................................................137
[:SOURce<hw>]:BB:GPS:TRIGger:ARM:EXECute
Stops signal generation for trigger modes "Armed_Auto" and "Armed_Retrigger". A sub-
sequent internal or external trigger event restarts signal generation.
Example:
BB:GPS:TRIG:SOUR INT
sets internal triggering.
BB:GPS:TRIG:SEQ ARET
sets Armed_Retrigger mode, i.e. every trigger event causes signal
generation to restart.
BB:GPS:TRIG:EXEC
executes a trigger, signal generation is started.
BB:GPS:TRIG:ARM:EXEC
signal generation is stopped.
BB:GPS:TRIG:EXEC
executes a trigger, signal generation is started again.
Usage: Event
[:SOURce<hw>]:BB:GPS:TRIGger:EXECute
Executes a trigger. The internal trigger source must be selected using the com-
mand :BB:GPS:TRIG:SOUR INT and a trigger mode other than AUTO must be selected
using the command :BB:GPS:TRIG:SEQ.