Specifications

User InterfaceGPS and Assisted GPS
40Operating Manual 1171.5248.12 ─ 15
Configuration of eight satellites is only possible for two path instruments and enabled by
parameter "Use Baseband A+B".
Adjust Total Power to 0dB
(R&S SMx/AMU instruments only)
Sets the power level of each satellite so that the sum of all levels results in 0 dB relative
to the global level setting of the instrument.
The power ratio among the individual satellites is maintained.
SCPI command:
[:SOURce<hw>]:BB:GPS:POWer:ADJust on page 93
Total Power
(R&S SMx/AMU instruments only)
Displays the total power of all satellites.
The total power is calculated from the power ratio of the activated satellites.
After "Adjust Total Power to 0dB", this power corresponds to 0 dB.
If the value is not equal to 0 dB, the individual activated satellites (whilst still retaining the
power ratios) are internally adapted so that the "Total Power" for achieving the set output
level is 0 dB.
SCPI command:
[:SOURce<hw>]:BB:GPS:POWer[:TOTal] on page 93
Use Spreading
(enabled for Generic mode only)
Satellite Configuration