Technical data
28 
6.  INSTALLATION INSTRUCTIONS 
  Caution: 
  The  sterilizer  must  be  placed  on  a  rigid  and  leveled  surface.  The  counter 
must be able to hold the load of the device and loaded material as follows: 
Model 
EZ9  EZ10, EZ10K 
lb 
106  117 
Load on Counter 
kg 
48  53 
Note: 
Make sure while placing the autoclave, to leave space around the machine, 
to give the technician access to service the machine. 
6.1  Electrical 
  The  electrical  connection  should  comply  with  the  device's  power 
requirement. It must also comply with local installation and safety rules 
and regulations. The voltage supplied to the device must comply with 
the label ± 5%. 
  In order to avoid any injury by electrical hazard, it is mandatory for the 
customer to have installed an earth leakage relay (GFI outlet or circuit 
breaker)  in  the  electrical  circuit  to  which  the  autoclave  is  connected. 
This relay disconnects all the poles of the electrical power line in case 
of  accidental  contact  with  the  autoclave’s  metal  enclosure,  by  the 
operator or another person, leading to a dangerous leakage current. 
Note:  Keep  the  back  and  the  right  side  of  the  autoclave  approximately  1” 
(25mm) away from the wall to allow for ventilation. 
  Connect the power cord to the socket on the rear side of the autoclave; 
plug  it  into  the  supply  outlet.  The  autoclave  must  be  connected  to  a 
properly grounded outlet. 
6.1.1  Setup 
  Your new  Tuttnauer Autoclave was  set at  the  factory and 
requires a minimal of setup. 
♦  Make sure the counter is level and sturdy. 
♦  Make sure all the feet are on the autoclave and none have 
been lost. 
♦  Position the autoclave on the counter. 
♦  Fill the reservoir with distilled water (see sec 6.5). 
♦  The unit is ready to operate. 