Technical data
7 
2.3  Environmental Emission Information 
1.  The peak sound level generated by the autoclave is less than 70 dBA 
with background noise of 60 dBA. 
2.  The total heat per hour transmitted by the autoclave is < 100 Wh for 
all models. 
2.4  Operating Conditions 
  This device is to be used for indoor use. 
This  autoclave  is  intended  for  NORMAL  environment  conditions  as 
follows: 
♦  Altitude up to 2000m. 
♦  Room temperature range 5ºC to 40ºC. 
♦  Installation Category II. 
♦  Pollution Degree 2. 
♦  Maximum  relative  humidity  80%  for  temperature  up  to  31ºC 
decreasing linearly to 50% relative humidity at 40ºC. 
♦  Mains
supply
voltage
fluctuations
up
to
+/-10%
of
the
nominal voltage. 
  The sterilizer should be loaded only with autoclavable material. 
  Caution! 
  Waste water should be brought into the public net in accordance with 
the local rules or requirements, i.e. 
  ONLY  NON-HAZARDOUS  LIQUIDS  SHALL  BE  DISPOSED  IN 
PUBLIC SEWAGE! 
2.5  Utilities 
Utilities  Unit  Value 
V-A  1ph, 120V – 16A,50/60 Hz 
Power supply (as appropriate) 
V-A  1ph, 230V – 16A,50/60 Hz 
Attention: 
The electrical net must be protected with a current leakage safety relay. 
The electrical network must comply with local rules or regulations. 
2.6  Construction 
  The  main  parts  of  the  autoclave  are  made  of  materials  as  indicated 
below: 
♦  Chamber is electro-polish and built of stainless steel 316 L. 
♦  Door is made of stainless steel CF8. 
♦  Trays are made of stainless steel 304. 
♦  Water reservoir is made of hard plastic material. 
♦  Door handle  is  made  of  hard  plastic  material, which is safe  to touch 
and thermo-insulated. 
♦  Covers are made of aluminum sheet, coated with Epoxy paint. 