8.3.4 Taper calibration
Saves the calibration
Confirm
Goes back to the reserved setup screen
Return to main screen
Open the spindle and fit the calibration tool fastening it to the adapter with the screw and nut provided. 1.
Position the tool in such a way that the laser crosses its central part during normal reading movement.2.
C3. lose the guard and press the button
When prompted on the screen, remove the tool and then fit it on the other side.4.
do N o T o p e N T h e s p I N d l e .
When reading is complete, remove the calibration tool.5.
dA N G e R ! do N o T p e R f o R m A s p I N W I T h T h e T o o l f I T T e d !
6. Save the calibration.
7. Press to go back to the main screen.