11.5. Visual acuity test results in overcorrection or undercorrection
① The test distance could be changed. Measure the test distance and adjust it to within the specification of this
device.
11.6. Red/Green test is overcorrection or undercorrection
① When it is possible to adjust the brightness of inspection room. Adjust it properly.
② Check the wavelength balance of red/green chart. The ambient illumination may affect the test result. In
setup menu adjust the red/green balance.
11.7. Red/Green filter is not filtering completely
① When Red/Green chart color doesn’t match with the Red/Green filter glass, the opposite color may be seen.
Please in setup menu adjust Red/Green color
11.8. Polarizer test is not working
① If a patient is not in the front of the device or stand aside, polarized filter might be less effective. Please check
patient’s position