User's Manual

--------------------------------------------------------------------------------------------------------------------- CDC-1 User Manual
55
Manufacturer announcement – electromagnetic waves trouble
Electromagnetic waves trouble
CDC-1 should be used in the below mentioned electromagnetic wave environment. CDC-1 purchaser
or user needs to confirm whether CVS is used in this type of environment.
Trouble test
Question of appropriateness
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker IEC 61000-3-3
Complies
Manufacturer announcement – electromagnetic waves tolerance
electromagnetic waves tolerance
CVS is to be used in the below designated electromagnetic wave environment. CVS
customer and user need to guarantee that the CVS will be used in this type of environment.
Tolerance test
IEC 60601
test level
Appropriateness level
Electrostatic
discharge(ESD)
IEC 6100042
Contact-Direct ±8 kV
Contact-Indirect ±8 kV
Note1)*
direct ±8 kV
indirect ±8 kV
Electric rapid
transients/bust
IEC 6100044
power supplying line ±2 kV
input/output line ±1 kV
power supplying line ±2 kV
input/output line ±1 kV
Surge
IEC 6100045
between lines ±1 kV
between line and grounding ±2 kV
differential mode ±1 kV
common mode ±2 kV
Voltage dip, instantaneous
interruption, voltage
fluctuation at the power
input line
IEC 61000411
For 0.5 cycle < 5 %UT(UT’s >
95 % decrease)
For 5 cycle 40 %UT(UT’s 60 %
decrease)
For 25 cycle 70 %UT(UT’s 30 %
decrease)
For 5 seconds < 5 %
UT(UT’s > 95 % decrease)
For 0.5 cycle < 5 %UT(UT’s > 95 %
decrease)
For 5 cycle, 40 %
UT(UT’s 60 % decrease)
For 25 cycle, 70 %UT(UT’s 30 %
decrease)
For 5 seconds, < 5 %
UT(UT’s > 95 % decrease)
Power frequency magnetic
field
(50/60 Hz)
IEC 6100048
30 A/m
30 A/m
Other UT is the a.c. power voltage for before approving the test level.
Note1)* ESD Test will be conducted with screw, lever, signal ports, enclosure, display, measurement
button and power switch including VCP/HCP as criteria ‘A’. But, criteria of connecting external monitor
will be applied ‘B’ because flickering phenomenon of connecting external monitor during interference of
electrostatic discharge doesn’t affect essential requirement.
14. EMC information