TRMS MULTIMETER E NG LI S H User’s manual C.
PRECAUTIONS FOR USE This instrument is compliant with the NF EN 61010-1 + NF EN 61010-2-030 safety standard for 1000 V voltages in category III or 600 V in category IV at an altitude below 2000 m, indoors, and with a maximum pollution level of 2. Failure to comply with safety instructions can create a risk of electric shock, fire, explosion and destruction of the instrument or the installations.
CONTENTS 1. OVERVIEW ......................................................................................................................................................... 4 1.1 The display ............................................................................................................................................ 4 1.2. The keys ................................................................................................................................................ 6 1.3.
1. OVERVIEW The C.
1.1.1 The display symbols Symbols Description AC Measurement of the AC signal DC Measurement of the DC signal AUTO Auto-ranging HOLD Memorisation and display of memorised values MAX Maximum RMS value MIN Minimum RMS value .run r.un ru.n Capacitance meter, acquisition in progress ----- Frequency measurement impossible O.
1.2. The keys The keyboard has 4 keys : MODE AC/DC, RANGE, MAX/MIN and HOLD. These are the keys: 1 2 3 4 Figure 2: the keyboard keys N°. Function 1 Selection of the display mode , , , ° C, ° F et du bar-graphe ( ) 2 Selection of the measurement scale and activation/deactivation of the screen backlighting ( 3 Activation of the MAX/MIN mode 4 Memorisation of the values and display mode Activation or deactivation of the automatic instrument shut-down 1.3.
1.4. The terminals These are the multimeter terminals: 1 2 3 Figure 4: the terminals N°. Input 1 6 A, 10 A current 2 Common 3 Other measurements 2. USE 2.1 First use Fit the battery shipped with the instrument as follows: 1. Using a screw driver, unscrew the four screws on the cover (item 1) located at the rear of the unit; 2. Place the battery in its housing (item 2) while respecting the polarity; 3. Screw the cover back on. 1 2 Figure 5: access to the battery 2.
2.4 The stand There are two stand positions, one for suspending the multimeter (position 1) and the other providing a stand (position 2). To change the stand position, proceed as follows: Position 1: fit the stand studs into the upper holes located on the rear of the unit: Position 2: fit the stand studs into the lower holes located on the rear of the unit: 3. FUNCTIONS 3.1 Switch functions To access the , , , , , or , position. Every position (except OFF) is confirmed by a beep.
3.1.2 Voltage measurement The instrument measures the following types of voltages: DC voltage with high impedance (DC); AC voltage with high impedance (AC); AC voltage with low impedance (VLowZ) In all cases "O.L " displays in excess of 1050 V and a beep sounds when the measurement exceeds 600 V. Volt VLowZ : This position is designed for measurements on electrical installations. The input impedance < 1 MΩ is used to avoid measuring "phantom" voltages caused by coupling between the lines.
3.1.4 Resistance measurement To measure the resistance, proceed as follows: 1. Set the switch to ; 2. Connect the black lead to the COM terminal and the red lead to "+"; 3. Place the test probes on the terminals of the component; Remark: all resistance measurements must be performed with the power off. 4. Read the measurement indicated on the display. 5. "O.L" is displayed if the circuit is open. 3.1.5 Audible continuity measurement For audible continuity measurement, proceed as follows: 1.
3.1.6 Diode test To measure and check a semiconductor junction, proceed as follows: 1. Set the switch to 2. Press twice on ; . The symbol is displayed; 3. Connect the black lead to the COM terminal and the red lead to "+"; 4. Place the test probes on the terminals of the component; 5. Read the junction threshold voltage measurement indicated on the display. 6. "O.L" is displayed if the circuit is open. 3.1.7 Capacitance measurement To measure the capacitance, proceed as follows: 1.
3.1.8 Temperature measurement To measure the temperature, proceed as follows: 1. Set the switch to 2. Press ; to select the unit and scale of the temperature (° C or ° F). Remark: the default measurement unit displayed is ° C. 3. Connect the temperature sensor to the COM and "+" terminals respecting the poles; 4. Read the measurement indicated on the display. 5. "O.L" is displayed the thermocouple is cut. NB: For greater accuracy, avoid subjecting the instrument to sudden temperature changes. 3.1.
Detection of fuse tripping - or meltdown: If the fuse has melted, the circuit between COM and the 10 A socket is cut. The display shows zero. 1. Set the switch to Ω. 2. Connect the V socket to the 10 A socket (see above); leave the "COM" socket free. 3. The display should show a result < 2 Ω, otherwise replace the fuse. 3.2 Key functions The functions: , , , can be accessed by pressing a key repeatedly, using short or long presses. The long-press function is shown by the pictogram under the key.
3.2.2 Key This key is used to manually select a measurement range or to activate the screen backlighting. The range defines the maximum range for measurements performed with the instrument. Remark: the Auto Range mode is activated by default. In normal mode Type of press short Resulting action on manually changes the measurement scale (extent and resolution). Example: In , mode, the screen displays: 59.00 V, range Auto When pressed the first time, the screen displays: 59.
In mode Type of press Resulting action short exit from on 15 mode.
3.2.3 Key This key displays the MAX, MIN modes. Max and Min fill in the highest and lowest values of the measurement. In normal mode Type of press Resulting action short enters the MAX/MIN mode. Note that in permanent operation. is displayed on to select the quantities for MAX or MIN Reminder: the MAX quantity is displayed by default. Example: display screen V / MAX. long (> 2 sec) quits the mode. on Remark: Auto power off is automatically activated. In + mode : refer to §.
3.2.4 Key This key is used to memorise the measurements and quantities or to deactivate the automatic power-off function on the instrument. In normal mode Type of press Resulting action short memorises a measurement status at a given moment and then makes it possible to view them successively on the display. on If activated, the Min Max acquisition continues as a background task. The bargraph continues to operate normally (even in Hold).
In mode Type of press Resulting action short on freezes the display of each MAX, MIN quantity displayed. MAX, MIN,acquisition continues in the background. This is indicated by blinking of the MAX, MIN symbols. If you press a second time, you quit the 18 mode.
4. SPECIFICATIONS 4.1 Reference conditions Influencing quantities: Reference conditions Temperature 23° C ± 5° C Relative humidity 45 % to 75 % Power supply voltage 9V±1V Frequency domain of the applied signal 40 Hz to 1 kHz Absence of electric field 4.2 Specifications of the reference conditions Uncertainties are given in: X % of the reading (R) ± Y counts (D). When the frequency exceeds 1 kHz, apply the formula given in the tables X % R + Y % x [F (kHz) - 0.4] R ± D.
4.2.2 AC voltages VLowZ AC position The bandwidth is reduced to 300 Hz - 3 dB. In VLowZ, and there is no 60 mV scale. Frequency measurement is performed in the same way as the voltage measurement: with a bandwidth of 300 Hz Additional uncertainty F(Hz) typ. Range Specified measurement range Resolution Uncertainty () 600 mV 60 - 600 mV 0.1 mV 1.2 % R + 5 D 6V 0.6 - 6 V 0.001 V 45 < F < 65 Hz : 0.3 % R 60 V 6 - 60 V 0.01 V at 100 Hz : 0.7 % R 1.2 % R + 3 D 600 V 60 - 600 V 0.
4.2.3 Frequency Special reference conditions : 150 mV < U < 600 V 0,15 A < I < 10 A When the switch is on the Hz or Volts position, the 300 Hz filter is not operational. When the switch is in the VLowZ position, the 300 Hz filter is activated for Volts and frequency Range 600 Hz 6 kHz Specified measurement range 10 - 600.0 Hz 0.01 - 6.00 kHz 0.01 - 10 kHz Uncertainty () 0.1 % R + 2 D 0.1 % R + 2 D 0.1 % R + 2 D 0.1 Hz 1 Hz Resolution 60 kHz 10 Hz - Below 10 Hz, the value is forced to zero.
4.2.
4.2.10 DC currents (10A DC) Range Resolution Specified measurement range Uncertainty (±) 6A 0.001 A 0.020 - 6.000 A 0.8 % R + 3 D Voltage drop Fast fuse 0.05 V / A 0.01 A 10 A / 20 A * 0.200 - 20.00 A Protection 0.8 % R + 2 D 10 A (or 11 A) /1000 V (*) Admissible overload: 10 A to 20 A for 30s max. with a 5 min pause between 2 measurements. Ambient temp. 35° C max. 4.2.11 AC currents (10A AC) Reminder: the measurement uncertainty is specified for a display of between 550 and 6000 points.
4.3 Operating conditions Operating conditions Temperature Relative humidity (RH) in use in storage -10° C to 50° C -20° C to 70° C 80 % RH at 50° C 90 % RH ( 45° C) 4.
4.6 Compliance with international standards Electrical safety Application of safety rules in compliance with standards NF EN 61010-1 + NF EN 61010-2-030 1000V, CAT III - 600V CAT IV, pollution level 2, double insulation.
4.7 Variations in the operating range Influencing quantity Battery voltage Range of Influence 7.5 V to 10 V Typical Influenced quantity influence MAX all 1D 0.2 % R + 1 D V DC mV 0.01 % R ± 0.2 D / 1° C 0.02 % R ± 0.25 D / 1° C VAC mV, VLowZ mV 0.08 % R ± 0.2 D / 1° C 0.15 % R ± 0.25 D / 1° C V DC 0.01 % R ± 0.1 D / 1° C 0.05 % R ± 0.1 D / 1° C VAC and VAC+DC Temperature 0.15 % R ± 0.1 D / 1° C A DC 0.05 % R ± 0.1 D / 1° C 0.1 % R ± 0.1 D / 1° C AAC and AAC+DC 0.08 % R ± 0.
5. MAINTENANCE For maintenance, only use the specified spare parts. 5.1 Cleaning Unplug all connections from the instrument and place the switch in the OFF position. Use a soft cloth slightly moistened with soapy water. Rinse with a damp cloth and dry quickly with a dry cloth or pulsed air. Make sure no foreign bodies hinder the operation of the cable fitting system. 5.2 Replacing the battery The symbol indicates that the battery is low.
6. WARRANTY Unless explicitly stated to the contrary, our warranty period is three years from the date the equipment is made available. Excerpt from our General Sales Terms and Conditions sent on request.
09 - 2018 X03706E02 - Ed. 2 DEUTSCHLAND - Chauvin Arnoux GmbH Straßburger Str. 34 - 77694 Kehl / Rhein Tel.: (07851) 99 26-0 - Fax: (07851) 99 26-60 ESPAÑA - Chauvin Arnoux Ibérica S.A C/ Roger de Flor N° 293, Planta 1- 08025 Barcelona Tel.: 902 20 22 26 - Fax: 934 591 443 ITALIA - Amra SPA Via Sant’Ambrogio, 23/25 - 20050 Bareggia di Macherio (MI) Tel.: 039 245 75 45 - Fax: 039 481 561 ÖSTERREICH - Chauvin Arnoux GmbH Slamastrasse 29 / 2 / 4 - 1230 Wien Tel.