Non-contractual document
906131102D - Ed 1
4.09 (1/2)
C100 series
Current clamp for AC current
DESCRIPTION
The C173 clamp measures leakage or differential currents from 1 mA upwards
and can also be used with multimeters equipped with a range in mV AC.
The C173 clamp measures earth-loop currents and leakage currents. It also
locates faults in circuits of single and three-phase networks.
For unearthed three-phase systems, use the optional Artifi cial Neutral.
ELECTRICAL SPECIFICATIONS
Current range:
0.001 AAC .. 1.2 AAC
0.01 AAC .. 12 AAC
0.1 AAC .. 120 AAC
1 AAC .. 1,200 AAC
Output signal:
1 V AC / AAC (1 V for 1 A)
100 mV AC / AAC (1 V for 10 A)
10 mV AC / AAC (1 V for 100 A)
1 mV AC / AAC (1 V for 1,000 A)
Accuracy and phase shift
(1)
:
1 A calibre
Primary current 0.001 A .. 0.01 A 0.01 A .. 0.1 A 0.1 A .. 1 A 1 A .. 1.2 A
% Accuracy of output signal 3 % + 1 mV 3 % + 1 mV 0.7 % + 1 mV 0.7 % + 1 mV
Phase shift not specified not specified 10° 10°
10 A calibre
Primary current 0.01 A .. 0.1 A 0.1 A .. 1 A 1 A .. 10 A 10 A .. 12 A
% Accuracy of output signal 1 % + 0.2 mV 0.5 % + 0.2 mV 0.5 % 0.5 %
Phase shift not specified 5 ° 2° 2°
100 A calibre
Primary current 0.1 A .. 1 A 1 A .. 10 A 10 A .. 100 A 100 A .. 120 A
% Accuracy of output signal 1 % + 0.2 mV 0.5 % + 0.2 mV 0.3 % 0.2 %
Phase shift not specified 2° 1° 1°
1,000 A calibre
Primary current 1 A .. 10 A 10 A .. 100 A 100 A .. 1,000 A 1,000 A .. 1,200 A
% Accuracy of output signal 1 % + 0.2 mV 0.5 % + 0.2 mV 0.2 % 0.2 %
Phase shift not specified 2° 1° 1°
Bandwidth:
10 Hz .. 3 kHz
Crest factor:
1 A calibre:
3 for I 3 A peak (1 A
RMS
)
10 A calibre:
3 for I 30 A peak (10 A
RMS
)
100 A calibre:
3 for I 300 A peak (100 A
RMS
)
1,000 A calibre:
3 for I 1700 A peak (500 A
RMS
)
Maximum currents:
1,000 A continuous for a frequency
500 Hz
(limitation proportional to the inverse of 1/2
of frequency beyond)
Load impedance:
≥ 10 MΩ and 47 pF
Output impedance:
1 A calibre: 10 kΩ ± 10 %
10 A calibre: 1 kΩ ± 10 %
100 A calibre: 100 Ω ± 10 %
1,000 A calibre: 100 Ω ± 10 %
Operating voltage:
600V
RMS
Common mode voltage:
600 V category III and pollution degree 2
Infl uence of adjacent conductor:
1 mA / A at 50 Hz
Infl uence of conductor position in jaws:
0.3 % of output signal for frequencies 400 Hz
Infl uence of frequency
(2)
:
1 A calibre:
<2 % of output signal 30 Hz .. 48Hz
and 65Hz .. 1 kHz
<10 % of output signal 1 kHz .. 3 kHz
10 A calibre:
<2 % of output signal 10 Hz .. 48Hz
and 65Hz .. 3 kHz
100 A calibre:
<1.5 % of output signal 10 Hz .. 48Hz
and 65Hz .. 3 kHz
1,000 A calibre:
<1 % of output signal 10 Hz .. 48Hz and 65Hz
.. 1 kHz
Infl uence of crest factor:
0.5 % for crest factor limited to 3
Infl uence of DC current superimposed on rated
current:
10 % at 1,000 A for a current DC from 10 A
Current
1 A 10 A 100 A 1,000 A
Output
1 V / A 100 mV / A 10 mV / A 1 mV / A
Current clamp for AC current
Model C173
(probe for leakage currents)