Pinces ampèremetriques pour courant AC
Non-contractual document
906131102D - Ed 1
4.08 (1/6)
C100 series
DESCRIPTION
This 1,000 A AC clamp can be used for easy display and measurement
of current curves. Equipped with a coaxial cable terminated by a BNC
connector, it is ideal for use with any oscilloscope. It outputs a signal in
mV directly proportional to the current. It offers 3 different sensitivities.
ELECTRICAL SPECIFICATIONS
Current range:
0.1 A AC .. 10 A AC (30 A peak)
1 A AC .. 100 A AC (300 A peak)
1 A AC .. 1,000 A AC (2,000 A peak)
Output signal:
100 mV AC / AAC (1 V for 10 A)
10 mV AC / AAC (1 V for 100 A)
1 mAAC / AAC (1 V for 1,000 A)
Accuracy and phase shift
(1)
:
10 A calibre
Primary current 0.1 A .. 0.5 A 0.5 A .. 2 A 2 A .. 10 A 10 A .. 12 A
% Accuracy of output
signal
3 % + 10 mV 3 % + 10 mV 3 % + 10 mV 3 % + 10 mV
Phase shift not specified not specified 15 ° 15 °
100 A calibre
Primary current 0.1 A .. 5 A 5 A .. 20 A 20 A .. 100 A 100 A .. 120 A
% Accuracy of output
signal
2 % + 5 mV 2 % + 5 mV 2 % + 5 mV 2 % + 5 mV
Phase shift not specified 15 ° 10° 5 °
1,000 A calibre
Primary current 1 A .. 50 A 50 A .. 200 A 200 A .. 1,000 A 1,000 A .. 1,200 A
% Accuracy of output
signal
1 % + 1 mV 1 % + 1 mV 1 % + 1 mV 1 % + 1 mV
Phase shift not specified 3° 2° 1°
Bandwidth:
10 Hz .. 100 kHz (-3 dB)
(depending on current value)
Rise/fall time from 10 % to 90 %:
3.5 µs
10 % delay time:
0.5 µs
Ampere second product:
10 A calibre: 3,2 A.s
100 A calibre: 26 A.s
1,000 A calibre: 64 A.s
Maximum currents:
1,000 A permanent
1,200 A for 40 minutes max. / > 20 minutes shutdown
for a frequency ≤ 1 kHz (limitation proportional to the
inverse of one third of the frequency beyond that)
Insertion impedance (at 400 Hz / 10 kHz)
10 A calibre: < 0.3 mΩ / < 6,6 mΩ
100 A calibre: < 0.3 mΩ / < 2 mΩ
1,000 A calibre: < 0.3 mΩ / < 1.6 mΩ
Output impedance at 1 kHz:
10 A calibre: 515 Ω ± 10 %
100 A calibre: 515 Ω ± 10 %
1,000 A calibre: 515 Ω ± 10 %
Infl uence of temperature:
150 ppm /k or 0.15 % of output signal per 10 °K
Infl uence of relative humidity:
< 0.1 % of output signal
Infl uence of adjacent conductor:
1mA / A at 50 Hz
Infl uence of DC current ≤ 30 A superimposed on
rated current:
< 1 %
Infl uence of conductor position in jaws:
0.1 % of output signal for frequencies 400 Hz
Infl uence of frequency
(2):
10 A calibre:
<10 % of output signal from 10 Hz .. 1 kHz
<5 % of output signal from 1 kHz .. 10 kHz
<20 % of output signal from 10 kHz .. 50 kHz
3 dB of output signal from 50 kHz .. 100 kHz
100 A calibre:
<5 % of output signal from 10 Hz .. 1 kHz
<3 % of output signal from 1 kHz .. 10 kHz
<20 % of output signal from 10 kHz .. 50 kHz
3 dB of output signal from 50 kHz .. 100 kHz
1,000 A calibre:
<1 % of output signal from 10 Hz .. 1 kHz
<2 % of output signal from 1 kHz .. 10 kHz
<10 % of output signal from 10 kHz .. 50 kHz
3 dB of output signal from 50 kHz .. 100 kHz
Infl uence of crest factor:
<1 % of output signal for crest factor 6 with
current
10 A calibre: 30 A peak
100 A calibre: 300 A peak
1,000 A
calibre: 3,000 A peak
Current
30 A peak 300 A peak 2,000 A peak
Output
100 mV / A 10 mV / A 1 mV / A
Oscilloscope clamp for AC current
Model C160
(insulated AC current probe)