Brochure
Table Of Contents
Serie MN
Pinces ampèremetriques pour courant AC
Non-contractual document
906131102D - Ed 1
4.08 (2/6)
C100 series
(1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48Hz à 1,000 Hz, distortion factor <1 % with no DC component,
external magnetic fi eld <40A/m, no DC components, no external conductor with circulating current, conductor centred for measurement,
load impedance: ≥ 1 MΩ and <100 pF
(2) Out of reference domain
MECHANICAL SPECIFICATIONS
•
Max. jaw opening:
53mm
•
Clamping capacity:
Cable: Ø max 52mm
Busbar: 1 busbar of 50 x 5mm/
4 busbars of 30x5mm
•
Operating temperature:
-10 °C to +55 °C
•
Storage temperature:
-40 °C to +70 °C
•
Relative humidity for operation:
0 to 85% RH decreasing linearly above 35 °C
•
Operating altitude:
0 to 2,000 m
•
Casing protection rating:
IP30 with clamp open (IEC 529)
IP40 with clamp closed (IEC 529)
•
Drop test:
1 m (IEC68-2-32)
•
Shock resistance:
100 g / 6 ms / half-periode (IEC 68-2-27)
•
Protection against impacts:
IK04 0.5 J (EN 50102)
•
Vibration resistance:
5/15Hz 1.5mm peak
15/25Hz 1mm peak
25/55Hz 0.25mm peak
(IEC68-2-6)
•
Self-extinguishing capability:
Casing and jaws: UL94 V0
•
Dimensions:
216 x 111 x 45mm
•
Weight:
550 g
•
Colours:
Dark grey case with red jaws
•
Output:
Via 2 m coaxial cable terminated by insulated
BNC plug
SAFETY SPECIFICATIONS
•
Electrical safety:
Instrument with double insulation or reinforced
insulation between the primary the secondary and
the grippable part located under the guard as per
IEC1010-1 & IEC1010-2-032
- 600V category III, pollution degree 2
- 300V category IV, pollution degree 2
•
Electromagnetic compatibility (EMC):
EN50081-1: classB
EN50082-2:
- Electrostatic discharge: IEC1000-4-2
without disturbance: 4 kV class 2
non-destructive: 15 kV class 4
- Radiated field: IEC1000-4-3
without disturbance: 10 V/m performance
criterionA
- Fast transients: IEC1000-4-4
without disturbance: 1 kV class 2
non-destructive: 2 kV class 3
- Magnetic field at 50/60 Hz: IEC1000-4-8
field of 400 A/m at 50 Hz: < 1 A
To order Reference
AC current clamp model C160 with operating manual P01120308
Oscilloscope clamp for AC current
Model C160
(insulated AC current probe)