DESCRIPTION
The D38N offers accurate AC current measurement
and a voltage output in mV allowing direct readings on
oscilloscopes. A switch with 3 positions on the handle
can be used to select the ranges. The wide opening of
the jaws means they can be used on cables and small
busbars.
ELECTRICAL SPECIFICATIONS
Current calibres:
1 AAC .. 30 AAC (90 A peak)
1 AAC .. 300 AAC (900 A peak)
1 AAC .. 2,400 AAC (9,000 A peak)
(3,000 A for temperature < 35 °C)
Output signal:
10 mV / AAC (3 V for 30 A)
1 mV / AAC (3 V for 300 A)
0.1 mV / AAC (3 V for 3,000 A)
Accuracy and phase shift
(1)
:
30 A calibre
Primary current 1.5 A 6 A 30 A 36 A
% Accuracy of
output signal
2 % ± 1 mV
Phase shift 20° 10° 5 ° 5 °
300 A calibre
Primary current 15 A 60 A 300 A 360 A
% Accuracy of
output signal
2 % ± 0.5 mV
Phase shift 1.5°
3,000 A calibre
Primary current 150 A 600 A 3,000 A 3600 A
% Accuracy of
output signal
2 % ± 0.2 mV
Phase shift 1.5°
Bandwidth:
10 Hz to 50 kHz (depending on current)
Rise/fall time from 10 % to 90 %:
4 µs
10 % delay time:
0.3 µs
Ampere second product:
30 A calibre: 30 A.s
300 A calibre: 125 A.s
3,000 A calibre: 180 A.s
Insertion impedance (at 400 Hz / 10 kHz):
30 A calibre: < 0.1 mΩ / < 1 mΩ
300 A calibre: < 0.1 mΩ / < 0.5 mΩ
3,000 A calibre: < 0.1 mΩ / < 0.4 mΩ
Maximum currents:
I < 2,400 A permanent
2,400 A .. 2,800 A for 10 minutes and then
30 minutes shutdown
2,800 A .. 4,000 A for 5 minutes and then
30 minutes shutdown
Output impedance:
30 A calibre: 130 Ω ± 15 %
300 A calibre: 140 Ω ± 15 %
3,000 A calibre: 140 Ω ± 15 %
Infl uence of temperature:
0.2 % of output signal par 10 K
Infl uence of adjacent conductor:
5 mA/A at 50 Hz
Infl uence of DC current < 10 % of rated calibre
superimposed on the rated current:
0.05 % / A DC
Infl uence of conductor position in jaws:
1 % + 0.1 A at 50/60 Hz
Infl uence of frequency
(2)
:
30 A calibre: < 1 dB from 10 Hz .. 10 kHz
300 A calibre: < 1 dB from 10 Hz .. 10 kHz
3,000 A calibre: < 1 dB from 10 Hz .. 10 kHz
MECHANICAL SPECIFICATIONS
Max. jaw opening:
90mm
Clamping capacity:
Cable: Ømax 64mm
Busbars:
5 busbars from 125 x 5mm
3 busbars from 100 x 10mm
(busbars spaced by their thickness)
Output:
Via 2 m coaxial cable terminated by insulated BNC
plug
Dimensions:
310 x 120 x 48mm
Weight:
1,200 g
Operating temperature:
-10 °C to +50 °C
Storage temperature:
-25 °C to +80 °C
Relative humidity for operation:
0 to 85% RH with a linear decrease above 35 °C
Operating altitude:
0 to 2,000 m
Casing protection rating:
IP 20 (IEC 529)
Drop test:
0.5 m (IEC 68-2-32)
Shock resistance:
100 g / 6 ms / half-periode (IEC 68-2-27)
Protection against impacts:
IK04 0.5 J (EN 50102)
Vibration resistance:
10/55/10 Hz, 0.15mm (IEC 68-2-6)
Self-extinguishing capability:
Handles: UL94 V0
Jaws: UL94 V2
Colours:
Dark grey handles with red jaws
SAFETY SPECIFICATIONS
Electrical safety:
Instrument with double insulation or reinforced
insulation between the primary the secondary and
the grippable part located under the guard as per
IEC 1010-1 & IEC 1010-2-032
- 600 V category III, pollution degree 2
- 300 V category IV, pollution degree 2
Current
90 A peak 900 A peak 9,000 A peak
Output
10 mV / A 1 mV / A 0.1 mV / A
Oscilloscope clamp for AC current
Model D38N
(insulated AC current probe)