User manual

SIMPLE LOGGER II English
Page 43
MODEL L111
Electrical
Input One
Input connection Two 4mm safety banana jacks
Input level 0 to 1A
AC
Uncertainties (50 / 60 Hz)
Resolution
0 to 10 mA: unspecified
10 to 50 mA: ± (0,5% of reading + 1 mA)
50 to 1000 mA: ± (0,5% of reading + 0,5 mA)
0.1 mA
Maximum input level 1,2 A
Input impedance
1
Sample rate 64 samples/cycle
Storage rate Programmable from 125 ms to 1 day
Storage modes Start/Stop, FIFO or Extended Recording Mode* (XRM™)
Recording length Programmable using DataView
®
from 15 minutes to 8 weeks
Memory 240,000 measurements (512 kB). The recorded data is stored in
non-volatile memory retained even if the battery is low or removed.
Communication USB 2.0 optically isolated
Power source ** 2 x 1.5 V AA (LR6) cell Alkaline batteries
Battery life 100 hours to >45 days (dependent on storage rate/recording length)
Mechanical
Dimensions 136x70x32 mm (5.38x2.75x1.28")
Weight (with batteries) 180 g (6.4 oz)
Case UL94-V0
Vibration IEC 68-2-6 (1.5 mm, 10 to 55 Hz)
Shock IEC 68-2-27 (30 G)
Drop IEC 68-2-32 (1 m)
Environmental
Operating temperature - 10°C to + 50°C (14°F to 122°F)
Storage temperature - 20°C to + 60°C (- 4°F to + 140°F)
Relative humidity up to 85% at 35°C (95°F) without condensation
Altitude 2000 m
Safety
Conformity EN 61010-1; 50 V CAT III; Pollution level 2
Level of protection IP 40
*See the end of chapter