5 SWITCH
Switch with 24 positions to select the functions and ranges.
(See drawing in § 12 - APPENDIX)
2 ANALOGUE DISPLAY
The dial comprises 6 scales:
■ 2 black scales, with antiparallax mirror,
for V DC and AC (0.100 and 0.30) and for A DC (0.100)
■ 1 green scale for Ω (0.10k)
■ 2 red scales for A AC with miniclamp MN 89 (0.3 and 0.10)
■ 1 red scale for dB (0.22)
1 TERMINALS
∅ 4 mm safety terminals
■ COM: common, terminal that receives the black lead
■ V
Ω Ω
Ω Ω
Ω : terminal that receives the red lead for voltages,
resistances and AC voltages with miniclamp MN 89
■■
■■
■ 10 A DC: terminal that receives the red lead for the 10 A DC ranges
■■
■■
■ mA DC: terminal that receives the red lead for the mA DC ranges
2 - DESCRIPTION
3 “FUS” LIGHT
Neon for testing the 1 A and 10 A fuses on current measurement.
If the «Fus.» light is lit, change the faulty fuse(s).
NB: This light requires the presence of a voltage
≥
80 V to come on.
4 “VOLTEST” LIGHT
Voltage presence light on ohmmeter.
If the «Voltest» light comes on: presence of a voltage
U > 15 V AC/DC. Unplug the leads and cut off the voltage
before making the resistance measurement.