User's Manual

Table Of Contents
Document Title
C-Scan System User Reference Manual
Document No. DOC0001667 Revision 14 Date 22 May 2022
Page 19 of 90
Before using this document, check that you use latest revision (see Arena)
* Printed revisions are for Reference Only*
Figure 2: Activation of the C-Scan
®
Track
4. Attach the Track to the subject's back:
Note: This step shall be performed only after activation of the C-Scan
®
Track.
Equipment needed: C-Scan
®
Track, C-Scan
®
Track position stencil, stickers, body
marker, razor and medical grade alcohol.
Note: The C-Scan
®
Track stickers must be applied directly to smooth skin for proper
adhesion. Make sure there are no obstacles, such as hair. If needed, shave the
designated attachment area. Clean the area with alcohol and let the skin airdry before
attaching the stickers.
The C-Scan
®
Track attachment process:
Prepare the subject for attaching the C-Scan
®
Track by marking its exact location on
the lower back using the provided stencil and body marker. as illustrated in Figure 3.
Figure 3: Illustration of the use of the stencil (location template)
Detailed steps:
Place the C-Scan
®
Track position stencil on the subject’s back and use the body
marker to mark the outline where the C-Scan
®
Track should be attached.
Figure 2A
Figure 2B