3.0 BASIC OPERATIONS
3.1 On/Off Switch
To turn the gauge on, press the ON/OFF Switch briefly
(see photo on page 3). Press and hold the ON/OFF Switch
for 2 seconds to turn the power off.
Note: The gauge will power off automatically after
90 seconds of non-use.
3.2 Selection of Measuring Units
1. Make sure the gauge is switched OFF.
2. Press and hold the ZERO and STATS keys simultaneously
3. While continuing to hold down the ZERO and STATS
keys, turn the power ON.
3.3 Manually setting the measuring mode
DCFN-3000EZ-E Only
In some cases, especially with varnish applied to zinc on steel,
it is advisable to manually set the measuring mode—F mode
(Ferrous) for measurements on steel/iron, N mode
(Non-ferrous) for measurements on non-ferrous metals.
Setting the measurement mode
1. Turn the gauge on by pressing the ON/OFF switch.
2. Press the ZERO and STATS keys simultaneously to
change the measuring mode. Each time the keys are
pressed the mode will change.
The display will show the mode selected (see photo,
page 4) using the following abbreviations:
12
7.0. SPECIFICATIONS
Range
DCF, DCFN Ferrous: 0.01 - 140.00 mils (0 - 3500 µm)
DCFN Only Non-Ferrous: 0.1 - 120.00 mils (3000 µm)
Accuracy ±0.1 mils (3µm) + 3% of reading
Resolution Refer To Section 8.0
Display Back-lit, 4-digit alphanumeric
Minimum
Measuring Area 0.4" x 0.4" (10mm x 10mm)
Minimum
Curvature Radius Concave: 2" (50mm) / Convex: 0.2" (5mm)
Minimum Substrate Thickness
DCF, DCFN Ferrous: 0.020 mils (0.5mm)
DCFN Only Non-Ferrous: 0.002 mils (0.05mm)
Calibration Factory calibration, user-set zero calibration,
Statistics Number of readings, mean value, standard
deviation, maximum and minimum reading
based on 99 readings (max).
Operating Temp. 32 °F to 122 °F (0 °C to 50 °C)
Surface Temp. 5 °F to 140 °F (-15 °C to 60 °C)
Storage Temp. -4 °F to 140 °F (-20 °C to +60 °C)
Battery 2x AAA (1.5V)
Dimensions 4" x 2" x 1" (107mm x 50mm x 25mm)
Weight 3.2 oz (90 g)
Protection Class IP 52 (proof against dust and dripping water)
Standards DIN, ISO, ASTM, BS
5
3.0 BASIC OPERATIONS
3.1 On/Off Switch
To turn the gauge on, press the ON/OFF Switch briefly
(see photo on page 3). Press and hold the ON/OFF Switch
for 2 seconds to turn the power off.
Note: The gauge will power off automatically after
90 seconds of non-use.
3.2 Selection of Measuring Units
1. Make sure the gauge is switched OFF.
2. Press and hold the ZERO and STATS keys simultaneously
3. While continuing to hold down the ZERO and STATS
keys, turn the power ON.
3.3 Manually setting the measuring mode
DCFN-3000EZ-E Only
In some cases, especially with varnish applied to zinc on steel,
it is advisable to manually set the measuring mode—F mode
(Ferrous) for measurements on steel/iron, N mode
(Non-ferrous) for measurements on non-ferrous metals.
Setting the measurement mode
1. Turn the gauge on by pressing the ON/OFF switch.
2. Press the ZERO and STATS keys simultaneously to
change the measuring mode. Each time the keys are
pressed the mode will change.
The display will show the mode selected (see photo,
page 4) using the following abbreviations:
12
7.0. SPECIFICATIONS
Range
DCF, DCFN Ferrous: 0.01 - 140.00 mils (0 - 3500 µm)
DCFN Only Non-Ferrous: 0.1 - 120.00 mils (3000 µm)
Accuracy ±0.1 mils (3µm) + 3% of reading
Resolution Refer To Section 8.0
Display Back-lit, 4-digit alphanumeric
Minimum
Measuring Area 0.4" x 0.4" (10mm x 10mm)
Minimum
Curvature Radius Concave: 2.0" (50mm) / Convex: 0.2" (5mm)
Minimum Substrate Thickness
DCF, DCFN Ferrous: 0.020 mils (0.5mm)
DCFN Only Non-Ferrous: 0.002 mils (0.05m)
Calibration Factory calibration, user-set zero calibration,
Statistics Number of readings, mean value, standard
deviation, maximum and minimum reading
based on 99 readings (max).
Operating Temp. 32 °F to 122 °F (0 °C to 50 °C)
Surface Temp. 5 °F to 140 °F (-15 °C to 60 °C)
Storage Temp. -4 °F to 140 °F (-20 °C to +60 °C)
Battery 2x AAA (1.5V)
Dimensions 4" x 2" x 1" (107mm x 50mm x 25mm)
Weight 3.2 oz (90 g)
Protection Class IP 52 (proof against dust and dripping water)
Standards DIN, ISO, ASTM, BS