Owner's manual
25
4.6 Connector of External input/output
4.6.1 Connector Pin Assign
External Input
Pin No. Signal Name Contents
1 +12V +12V power supply output
2 G12 Ground of power supply output
3 TRG_IN External pulse input
4 IN_COM Common of External pulse input
5 NC
6 NC
7 FG Frame ground
DT-329 side: SRCN2A16-7S [JAE]
Accessory: SRCN6A16-7P [JAE]
External Output
Pin No. Signal Name Contents
1 OUT External pulse output
2 OUT_COM Common of External pulse output
3 FG Frame ground
DT-329 side: SRCN2A13-3S [JAE]
Accessory: SRCN6A13-3P [JAE]
4.6.2 External Pulse Input
The unit can be operated and controlled by an external signal, from either a sensor or a machine signal like
an output from a printing press, re-winder or others. This will allow the stroboscope to flash in sync with the
target object’s rotational speed.
The input signal will have the following characteristics.
Input frequency: 60 - 10,000 FPM (1 – 160 Hz)
Input pulse width: over 50us
Delay angle: 0 - 359° (every 1° )
Delay time: 0 - 999ms (every 1ms)
Please connect with open collector output (below 1V (ON voltage)) or no-voltage contact output (approx.
11mA current)
Connector of
External
input
3pin
4pin
+12V
Inside DT-329
Photocoupler
1kΩ
[Input circuit]