4. The Dip Switch Block will be located 
at the top of the opening after 
removing the Slide Guide Plate.
5.32
 Setting Dip Switches
Using a small slotted screwdriver or other pointed tool, slide each of the five (5)
Dip Switches into the ON (up) or OFF (down) position as indicated in the illus-
tration below for the desired setting. Dip Switch #6 is not used.
Memory Mode   DIP-5  Data Stored in Memory
Standard Memory OFF  MIN, MAX, PEAK
Standard Memory 
Without Auto–  ON  MIN, MAX, PEAK
Power Off
Display Update Rate  DIP-1  DIP-2  DIP-3 DIP-4 
0.5 sec  ON  OFF OFF OFF
1.0 sec  OFF  ON OFF OFF
2.0 sec  OFF  OFF ON OFF
4.0 sec  OFF OFF OFF ON
123456
ON
DIP Switch #6 
(Not Used)
Factory switch:
DO NOT
TOUCH
DIP switches
14
6.00 BUILT-IN MEMORY SYSTEM
The DTMB Built-In Memory System can be configured to operate in one of two
modes: Standard Memory or Standard Memory With No Auto Power Off. The
data stored in the Standard Memory can be recalled to the display for viewing.
6.10 Standard Memory Mode
The DTMB takes 62 tension measurements per second and displays the aver-
age of these measurements over a user-selected Display Update Rate of 0.5, 1, 2
or 4 seconds. The Standard Memory system checks each displayed value and
retains the Maximum, Minimum and Peak tension values which occurred dur-
ing the Measuring Interval. The Measuring Interval is started and stopped by
pressing the key.
Maximum V
alue: highest 
displayed value during 
measuring interval 
Minimum
Value: lowest 
displayed value during 
measuring interval 
Peak V
alue: highest transient
valu of all of the measurements used to calculate each displayed value
Note: The “M” indicator on the LCD display will flash on and off, indicating
that the Standard Memory system is activated.
To Start/Stop measuring interval for Standard Memory, press  key.
STORE
STORE