Owner manual

Model ESM301 / ESM301L Version 2 Test Stand User’s Guide
15
Label Description
+ or –
Cycles through the available settings
ENTR
Returns to the Test Parameter Setup menu
ESC
Exits the parameter without saving changes
4.14 Programmable Button Function (KEYS)
Three button function modes are available:
1. Maintained
The crosshead will move continuously once the button has been pressed and held.
Subsequently pressing the UP, DOWN, or STOP during a test will stop crosshead motion.
2. Momentary
The crosshead will move only if the button is held down. Releasing the button will stop
movement immediately.
3. Auto
Holding down the button for more than 0.5 seconds will enter momentary mode, at which time
an audible indicator will sound and the LED indicator on the button pushed will be illuminated.
A short tap on the button will operate the test stand in maintained mode. Pressing UP,
DOWN, or STOP during maintained mode will stop crosshead motion.
Default setting: maintained
Available settings: maintained, momentary, auto
Label Description
+ or –
Cycles through the available settings
ENTR
Returns to the Test Parameter Setup menu
ESC
Exits the parameter without saving changes
4.15 Return to Default Settings (DEFAULT SETTINGS)
This setting provides a quick return to factory settings, as follows:
SPEED: 10 in/min
UP SPEED: 10 in/min
DOWN SP: 10 in/min
AUTO RETURN: off
CYCLING: 00000 (off)
HI LIMIT: +12.000
LO LIMIT: -12.000
OVERLD: off
PRELOAD: off
LOAD HOLDING: off
BREAK DETECT: off
CONTROL: CONSOLE
BAUD RATE: 115200
STOP & PAR: 8-1n
U N I T S : i n /
m
i n
E S C + E N T R
K E Y S : M A I N T A I N E D
E S C + E N T R