Owner manual

Model ESM301 / ESM301L Version 2 Test Stand User’s Guide
9
4.1 Speed, Up Speed, Down Speed (SPEED, UP SPEED, DOWN SP)
If the independent up and down speed option has not been enabled, the up and down speeds will be the
same, and is programmed in the SPEED parameter. If independent up and down speeds are enabled, UP
SPEED and DN SPEED parameters will be present, and may be set individually.
Default setting: 10 in/min
Available settings: 0.02 – 45 in/min
4.2 Auto Return (AUTO RETURN)
With this feature, the crosshead moves to a limit switch or soft limit (force set point, distance limit,
preload, or break detection), whichever occurs first, and stops. Then, the crosshead returns to the other
limit and stops. The test speed is dictated by the SPEED setting or the UP SPEED and DOWN SP
settings. The return speed is always maximum speed. The maximum speed depends on whether or not
the optional high speed range has been enabled.
Default setting: off
Available settings: off, on
Label Description
+ or –
Cycles through the available settings
ENTR
Returns to the Test Parameter Setup menu
ESC
Exits the parameter without saving changes
Note: If AUTO RETURN is turned on, CYCLING is automatically turned off and the KEYS parameter is
automatically set to MAINTAINED mode. See following pages for details on the CYCLING and KEYS
parameters.
U P S P E E D
:
1 0 . 7 3
E S C + E N T R
Label Description
+
Increments the speed setting by .01 in or .5 mm.
Holding down + will increment at an increasingly faster
rate.
Decrements the speed setting by .01 in or .5 mm.
Holding down will decrement at an increasingly
faster rate.
ENTR
Returns to the Test Parameter Setup menu
ESC
Exits the parameter without saving changes
A U T O R E T U R N o f f
E S C + E N T R