3.0 REMOVING AND REMOUNTING THE FILAMENT GUIDE
The tension meter is supplied with a FILAMENT GUIDE for fast and easy material 
acquisition. For application in hard-to-reach areas with limited access space, you can 
remove the FILAMENT GUIDE.
Removal Procedure
1  Loosen the SETSCREWS (2x) with the supplied screwdriver (1.5 mm blade).
2.  Unscrew and remove the GUIDE ROLLERS (2x) with the supplied open end 
wrench (jaw width 4 mm).
3.  Slip the FILAMENT GUIDE off the ROLLER SHAFTS.
4.  Screw the GUIDE ROLLERS (2x) back onto the ROLLER SHAFTS and carefully 
tighten using the supplied open end wrench (jaw width 4 mm) until hand-tight.
Remounting Procedure
1.  Unscrew and remove the GUIDE 
ROLLERS (2x) with the supplied 
open end wrench jaw width 4 mm).
2.  Slip the lament guide on to the 
ROLLER SHAFTS.
3.  Screw the GUIDE ROLLERS (2x) 
back on to the ROLLER SHAFTS 
and carefully tighten them using the supplied open end wrench (jaw width 4 mm) 
until hand-tight.
4.  Push the FILAMENT GUIDE forward far nough to ensure that the rollers do not rub 
against the FILAMENT GUIDE and that the process material can slide unhindered 
from the FILAMENT GUIDE into the roller grooves (g. CENTER LINE).
5.   Carefully tighten the SETSCREWS (2x) with the supplied screwdriver 
(blade width 1.5 mm) until hand-tight.
GUIDE ROLLERS (2x)
ROLLER SHAFTS (2x)
FILAMENT GUIDE
SETSCREWS (2x)
GRUB SCREWS (3x)
MEASURING 
ROLLER
CENTER LINE
– 15 –
o3rd calibration point:o
1. Repeat steps 1 through 4 from the preceding procedure using a weight which 
corresponds to 70% of the tension range
2.  Press the RECALL key. As long as the RECALL key is 
depressed, the DISPLAY shows a decimal value which is 
higher than the second decimal value, e.g., 8000. This decimal 
value may vary from instrument to instrument. Write down the 
value.
3. Release the RECALL key when the value shown on the 
DISPLAY is stable (the reading might uctuate greatly). 
The DISPLAY shows E 100
o4th calibration point:o
1. Repeat steps 1 through 4 from the preceding procedure using a weight which 
corresponds to 100% of the tension range
2.  Press the RECALL key. As long as the RECALL key is 
depressed, the DISPLAY shows a decimal value which is 
higher than the third decimal value, e.g., 9500. This decimal 
value may vary from instrument to instrument. Write down the 
value
3. Release the RECALL key when the value shown on the DISPLAY is stable (the 
reading might uctuate greatly).
4.  Press and then release the RECALL key. the DISPLAY switches off after 
approximately 1 second. The new calibration is stored. 
5. Press the POWER key. the instrument switches off.
6.  Verify the new calibration, following the directions in Sec. 7.0 If this procedure 
shows a deviation, you can recalibrate the tension meter again or restore the factory 
calibration as described in Sec. 8.3.
If the verication of the calibration according Sec.7.0 shows a deviation beyond 
the allowable tolerance and a reliable operation is no longer allowed, the 
instrument has to be returned to Electromatic for factory recalibration. 
8000
9500
E 100